FDA Adverse Event Injury Summary report: N

STYLE 115 SILICONE GEL FILLED BREAST IMPLANT

MDR report key: 2251798 · Received September 13, 2011

Report

Report Number
2024601-2011-00780
Event Type
Injury
Date Received
September 13, 2011
Date of Event
August 11, 2011
Report Date
August 22, 2011
Manufacturer
ALLERGAN
Product Code
FTR
PMA / PMN Number
P020056
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

MEDWATCH SUBMITTED ON (B)(4) 2011. DEVICE LABELING ADDRESSES THE REPORTED EVENT OF SEROMA AND CAPSULAR CONTRACTURE AS FOLLOWS: THE CORE STUDY: 7 YEAR ADVERSE EVENT RATES: FOR PRIMARY AUGMENTATION PATIENTS, SEROMA RATE = 1.6 PERCENT. CAPSULAR CONTRACTURE 15.5 PERCENT. SWELLING = 7.1 PERCENT. "AFTER BREAST IMPLANT SURGERY THE FOLLOWING MAY OCCUR AND/OR PERSIST, WITH VARYING INTENSITY AND/OR FOR A VARYING LENGTH OF TIME: HEMATOMA/SEROMA..." (ALLERGAN SILICONE LABELING). DEVICE LABELING REVIEWED: THERE WERE NO REPORTED EVENTS OF LYMPHOMA/ALCL, FOR PATIENTS IN THE CORE STUDY, IN THE LABELING FOR SILICONE IMPLANTS.

Description of Event or Problem · 1

PATIENT REPORTS VIA EMAIL A CASE OF LYMPHOMA. SHE REPORTS THAT SHE IS AN AUGMENTATION PATIENT AND WAS IMPLANTED IN 2007. SHE STATES THAT SHE NOTICED HER IMPLANT GETTING HARD AND WENT TO SEE THE DOCTOR. ON (B)(6) 2011 THE SURGEON CALLS IN TO PROVIDE ADDITIONAL INFORMATION ON THIS PATIENT. HE PRESENTS HER HISTORY AS A RECONSTRUCTION PATIENT. SHE WAS DIAGNOSED WITH RIGHT BREAST CANCER PRIOR TO IMPLANTATION. RECONSTRUCTION SURGERY TOOK PLACE ON (B)(6) 2007, WITH THE LEFT SIDE BEING DONE FOR PROPHYLACTIC PURPOSES. ON (B)(6) 2011 THE PRESENTED WITH CAPSULAR CONTRACTURE BAKER GRADE IV OF THE RIGHT BREAST AND WAS EXPLANTED. THE DIAGNOSIS OF ALCL WAS MADE FROM THE RIGHT BREAST CAPSULE SAMPLE FROM THE (B)(6) 2011 SURGERY. THE PATIENT IS SCHEDULED TO HAVE THE DEVICES REMOVED BILATERALLY (B)(6) 2011.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 STYLE 115 SILICONE GEL FILLED BREAST IMPLANT FTR ALLERGAN NA NI

Patients

Seq Age Sex Outcome Treatment
1 60 YR Required Intervention