FDA Adverse Event Injury Summary report: N

MINICAP TRANSFSET

MDR report key: 2251652 · Received September 16, 2011

Report

Report Number
1423500-2011-12327
Event Type
Injury
Date Received
September 16, 2011
Date of Event
August 1, 2011
Report Date
August 30, 2011
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
PMA / PMN Number
K882498
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A BATCH REVIEW WAS CONDUCTED FOR POTENTIALLY ASSOCIATED LOT NUMBERS H10K05047, H11A13057 AND H11C24092, H11D25048 AND NO EXCEPTIONS WERE OBSERVED THAT WERE RELATED TO THE REPORTED CONDITION. THE CAUSE OF THE USER ERROR WHICH RESULTED IN THE PERITONITIS WAS UNDETERMINED. THE LABEL REVIEW FOUND THE LABELING ADEQUATE FOR THE USER ERROR IN THIS COMPLAINT. BAXTER HAS CONDUCTED A TREND REVIEW AND FOUND THAT SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.

Additional Manufacturer Narrative · 1

(B)(4). AS THE DATE OF ONSET OF THIS PERITONITIS EPISODE IS UNKNOWN AND PATIENTS DISCARD SUPPLIES AFTER EACH THERAPY, THE SAMPLE WAS NOT REQUEST. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, A FOLLOW-UP REPORT WILL BE SUBMITTED. THIS IS REPORT 3 OF 4 INVOLVED IN THIS PERITONITIS EVENT.

Description of Event or Problem · 1

THE FOLLOWING INFORMATION WAS RECEIVED FROM GLOBAL PHARMACOVIGILANCE (GPV) AND IS A SPONTANEOUS REPORT BY A CONSUMER IN THE USA OF PERITONITIS IN A PATIENT COINCIDENT WITH DIANEAL PD4 AMBUFLEX THERAPY. DURING A CALL WITH BAXTER CUSTOMER SERVICES, THE FOLLOWING WAS REPORTED. ON AN UNREPORTED DATE, THE PATIENT MADE A MISTAKE/TOUCH CONTAMINATION. ON (B)(6) 2011, THE PATIENT EXPERIENCED PERITONITIS. ON (B)(6) 2011, THE PATIENT WAS HOSPITALIZED FOR THE PERITONITIS. THE CAUSE OF THE PERITONITIS WAS REPORTED AS PATIENT MADE A MISTAKE/TOUCH CONTAMINATION. TREATMENT INFORMATION WAS NOT PROVIDED. ON (B)(6) 2011 THE PATIENT WAS DISCHARGED AND WAS RECOVERING FROM THE PERITONITIS. THE OUTCOME FOR THE PATIENT MADE A MISTAKE/TOUCH CONTAMINATION WAS NOT REPORTED. DIANEAL PD4 AMBUFLEX THERAPY WAS ONGOING. AN OPINION OF CAUSALITY WAS NOT PROVIDED. THE NURSE DECLINED TO PROVIDE FURTHER INFORMATION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 MINICAP TRANSFSET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME

Patients

Seq Age Sex Outcome Treatment
1 77 YR Hospitalization DIANEAL PD4 AMBUFLEX