FLOCARE INFINITY 2 PUMP

MDR report key: 2251605 · Received August 30, 2011

Report

Report Number
1722139-2011-00282
Event Type
Malfunction
Date Received
August 30, 2011
Date of Event
December 3, 2009
Report Date
December 21, 2009
Manufacturer
MOOG MEDICAL DEVICES GROUP
Product Code
LZH
PMA / PMN Number
K031199
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

NO EVIDENCE OF FIRE WAS FOUND. THE PCB WAS CHARRED AROUND A DIODE. THE AC ADAPTOR WAS RETURNED TO THE MFR FOR ANALYSIS WHO FOUND A LOOSE CONNECTION BETWEEN A RESISTOR AND SOLDER PAD, WHICH MAY BE A RESULT OF MECHANICAL STRESS AFTER MANY YEARS IN THE FIELD. THE CONSTANT HIGH VOLTAGE ON THESE PARTS CAUSED EXCESSIVE POWER DISSIPATION BY THE DIODES WHICH RESULTED IN OVERHEATING, BURNING THE PUMP PCB AND MELTING THE KEYPAD. THIS MDR IS BEING FILED AS PART OF THE RETROSPECTIVE REVIEW FOR REPORTABILITY.

Description of Event or Problem · 1

THE CUSTOMER ALLEGED THE PUMP CAUGHT FIRE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FLOCARE INFINITY 2 PUMP LZH MOOG MEDICAL DEVICES GROUP INFINITY PUMP

Patients

Seq Age Sex Outcome Treatment
1 66 YR Disability