FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2251442 · Received September 9, 2011

Report

Report Number
1720753-2011-21792
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
August 31, 2011
Report Date
September 9, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE SVC REP REPLACED THE VIDEO CONTROLLER BOARD. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SVC.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYS DISPLAYED OVER EXPOSED IMAGES ON THE MONITOR. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1