FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 2251442
·
Received September 9, 2011
Report
- Report Number
- 1720753-2011-21792
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Date of Event
- August 31, 2011
- Report Date
- September 9, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TX, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ONSITE INVESTIGATION. THE SVC REP REPLACED THE VIDEO CONTROLLER BOARD. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SVC.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYS DISPLAYED OVER EXPOSED IMAGES ON THE MONITOR. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |