FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2251431 · Received September 9, 2011

Report

Report Number
1720753-2011-21784
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
September 1, 2011
Report Date
September 9, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REPRESENTATIVE FLASHED THE NODES, REFORMATTED THE HARD DRIVE, AND RELOADED THE SOFTWARE. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT DISPLAY THE IMAGE DURING ROTATION. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 9800

Patients

Seq Age Sex Outcome Treatment
1