FDA Adverse Event Malfunction Summary report: N

9800

MDR report key: 2251398 · Received September 9, 2011

Report

Report Number
1720753-2011-21791
Event Type
Malfunction
Date Received
September 9, 2011
Date of Event
August 23, 2011
Report Date
September 9, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC0
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SD, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REPRESENTATIVE REPLACED THE GENERATOR DRIVER BOARD. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A CHANNEL 15 ERROR MESSAGE AND WOULD NOT PERFORM FLUOROSCOPY. THIS HAPPENED DURING A CASE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 9800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC0 9800

Patients

Seq Age Sex Outcome Treatment
1