FDA Adverse Event
Malfunction
Summary report: N
9800
MDR report key: 2251398
·
Received September 9, 2011
Report
- Report Number
- 1720753-2011-21791
- Event Type
- Malfunction
- Date Received
- September 9, 2011
- Date of Event
- August 23, 2011
- Report Date
- September 9, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC0
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SD, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SERVICE REPRESENTATIVE PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REPRESENTATIVE REPLACED THE GENERATOR DRIVER BOARD. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYSTEM DISPLAYED A CHANNEL 15 ERROR MESSAGE AND WOULD NOT PERFORM FLUOROSCOPY. THIS HAPPENED DURING A CASE. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9800 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC0 | 9800 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |