FDA Adverse Event Malfunction Summary report: N

S3 SHOULDER PLATE 3 HOLE LEFT

MDR report key: 2251032 · Received August 26, 2011

Report

Report Number
1818910-2011-16136
Event Type
Malfunction
Date Received
August 26, 2011
Report Date
August 1, 2011
Manufacturer
DEPUY ACE S.A.'R.L.
Product Code
LXT
PMA / PMN Number
K060290
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
FL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE DEVICES ASSOCIATED WITH THIS REPORT WERE NOT RETURNED. ALTHOUGH THE COMPLAINT IS NON-VERIFIABLE, THERE ARE PREVIOUS COMPLAINTS FOR THIS FAILURE MODE. PATIENT ANATOMY AND SURGICAL TECHNIQUE CAN IMPACT THE PEG ENGAGEMENT EVEN IF THE PRODUCT IS WITHIN SPECIFICATION. THIS IS DISCUSSED IN DVA-104443-DFMEA AND DVA-104443-RMR AND IS APPROVED WITH AN IFR (INVESTIGATE FURTHER REDUCTION) DESIGNATION. THE ENGAGEMENT ISSUES ARE IMPACTED BY PATIENT ANATOMY AND SURGICAL TECHNIQUE. ALTHOUGH THE RISK IS CONSIDERED ACCEPTABLE PER THE RMR (RISK MANAGEMENT REPORT), PRODUCT DEVELOPMENT, QUALITY AND MARKETING ARE INVESTIGATING WAYS TO FURTHER REDUCE THE OCCURRENCE OF PEG ENGAGEMENT ISSUES. A SEARCH OF THE COMPLAINT DATABASE FOUND NO PRIOR REPORTS FOR BOTH OF THE REPORTED PART AND LOT NUMBER COMBINATIONS. (B)(4) WAS OPENED ON (B)(4) 2011 TO DETERMINE IF ANY ADDITIONAL ACTIONS MAY DECREASE THE LIKELIHOOD OF PEG ENGAGEMENT ISSUES. ANY ROOT CAUSE AND/OR CORRECTIVE ACTIONS WILL BE DOCUMENTED IN (B)(4). DEPUY CONSIDERS THE INVESTIGATION CLOSED. SHOULD ANY ADDITIONAL INFORMATION BE RECEIVED TO CHANGE THE OUTCOME OF THE PERFORMED INVESTIGATION, THE COMPLAINT WILL BE RE-OPENED.

Additional Manufacturer Narrative · 1

THIS COMPLAINT IS STILL UNDER INVESTIGATION. DEPUY WILL NOTIFY THE FDA OF THE RESULTS OF THIS INVESTIGATION ONCE IT HAS BEEN COMPLETED.

Description of Event or Problem · 1

THE PEGS WOULD NOT ENGAGE THE THREADED HOLES IN THE PLATE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 S3 SHOULDER PLATE 3 HOLE LEFT 87LXT LXT DEPUY ACE S.A.'R.L. NA RM051A

Patients

Seq Age Sex Outcome Treatment
1 76 YR