GORE EXCLUDER AAA ENDOPROSTHESIS
Report
- Report Number
- 2953161-2011-00194
- Event Type
- Injury
- Date Received
- September 12, 2011
- Date of Event
- August 16, 2011
- Report Date
- September 13, 2011
- Manufacturer
- W.L. GORE & ASSOCIATES
- Product Code
- MIH
- PMA / PMN Number
- P020004
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- PHYSICIAN
Narratives
A REVIEW OF THE MANUFACTURING PAPERWORK HAS BEEN CONDUCTED. THE REVIEW OF THE MANUFACTURING PAPERWORK VERIFIED THAT THIS LOT MET ALL PRE-RELEASE SPECIFICATIONS. ACCORDING TO THE GORE EXCLUDER AAA ENDOPROSTHESIS INSTRUCTIONS FOR USE THE (B)(4) IS INTENDED TO TREAT PTS WITH AN AORTIC DIAMETER OF 24-26MM.
ON (B)(6) 2011 THE PT WAS BEING TREATED WITH THE GORE EXCLUDER AAA ENDOPROSTHESIS. THE FINAL ANGIOGRAM SHOWED A PROXIMAL TYPE I ENDOLEAK DUE TO POOR WALL APPOSITION SO THE PHYSICIAN USED A COOK CODA BALLOON TO MODEL THE PROXIMAL END OF THE TRUNK IPSILATERAL LEG COMPONENT. AFTER THE BALLOONING THE PTS BLOOD PRESSURE DROPPED, BUT STABILIZED. LATER THAT DAY A CT SHOWED A DISSECTION PROXIMAL TO THE TRUNK IPSILATERAL LEG COMPONENT AND THE PT WAS CONVERTED TO OPEN SURGICAL REPAIR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | GORE EXCLUDER AAA ENDOPROSTHESIS | MIH / SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT | MIH | W.L. GORE & ASSOCIATES | WLG425 | 8678443 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 78 YR | Hospitalization| O| R | PIGTAIL CATHETER| COOK CODA BALLOON| TERUMO WIRE| LUNDERQUIST GUIDEWIRE |