FDA Adverse Event
Malfunction
Summary report: N
ESPRIT VENTILATOR
MDR report key: 2250963
·
Received September 16, 2011
Report
- Report Number
- 2031642-2011-00323
- Event Type
- Malfunction
- Date Received
- September 16, 2011
- Date of Event
- August 26, 2011
- Report Date
- August 26, 2011
- Manufacturer
- RESPIRONICS CALIFORNIA, INC.
- Product Code
- CBK
- PMA / PMN Number
- K981072
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AK, US
- Reporter Occupation
- OTHER
Narratives
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE UNIT WOULD NOT POWER ON. THE VENTILATOR WAS NOT IN USE THEREFORE THERE WAS NO PATIENT INVOLVEMENT OR HARM. THE SERVICE TECHNICIAN REPLACED THE POWER SUPPLY TO ADDRESS THE FINDING. PERFORMANCE VERIFICATION TESTING WAS COMPLETED AND PASSED TO OPERATING SPECIFICATIONS. ANALYSIS OF THE POWER SUPPLY REVEALED A MALFUNCTION THAT MAY RESULT IN THE UNIT SHUTTING DOWN DURING OPERATION WITH AN ACTIVE BACKUP ALARM WHEN AC POWER IS RESTORED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ESPRIT VENTILATOR | VENTILATOR, CONTINUOUS | CBK | RESPIRONICS CALIFORNIA, INC. | V1000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |