PRISMAFLEX SETS (OXIRIS)
Report
- Report Number
- 8010182-2025-00393
- Event Type
- Malfunction
- Date Received
- July 15, 2025
- Date of Event
- June 21, 2025
- Report Date
- August 13, 2025
- Manufacturer
- VANTIVE US HEALTHCARE LLC
- Product Code
- KDI
- UDI-DI
- 07332414124090
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CH
- Reporter Occupation
- NURSE
- Health Professional
- Yes
Narratives
ADDITIONAL INFORMATION: H6, H11. H11: THE ACTUAL DEVICE WAS NOT AVAILABLE; HOWEVER, PHOTOGRAPHS OF THE SAMPLE WERE PROVIDED FOR EVALUATION. VISUAL INSPECTION OF THE PHOTOS SHOWED A LEAK FROM THE SET ACCESS PRE-PUMP POD. THIS COMPONENT IS MANUFACTURED BY A VANTIVE EXTERNAL SUPPLIER. A BATCH REVIEW WAS CONDUCTED AND THERE WERE NO DEVIATIONS FOUND RELATED TO THIS REPORTED CONDITION DURING THE MANUFACTURE OF THIS LOT. THE REPORTED CONDITION WAS VERIFIED. THE CAUSE OF THE POD DEFECT WAS DETERMINED TO BE RELATED TO SUPPLIER MANUFACTURING. TWO (2) NONCONFORMANCE RECORDS HAVE BEEN OPENED TO INVESTIGATE AND ADDRESS THIS ISSUE. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
OXIRIS C IS SIMILAR TO OXIRIS AND OXIRIS S. OXIRIS AND OXIRIS S HAVE BEEN TEMPORARILY APPROVED FOR USE IN THE US UNDER EMERGENCY USE AUTHORIZATION EUA: (B)(4) WITH A SPECIFIC INDICATION TO TREAT PATIENTS WITH COVID-19 INFECTION. SHOULD ADDITIONAL RELEVANT INFORMATION BECOME AVAILABLE, A SUPPLEMENTAL REPORT WILL BE SUBMITTED.
IT WAS REPORTED THAT AFTER 10 HOURS OF CONTINUOUS RENAL REPLACEMENT THERAPY (CRRT) USING A PRISMAFLEX OXIRIS FILTER SET, THE PRESSURE SENSOR MEMBRANE WAS FOUND DAMAGED AND LEAKING BLOOD. THERE WAS NO REPORT OF PATIENT INJURY OR MEDICAL INTERVENTION ASSOCIATED WITH THIS EVENT. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 159858 | PRISMAFLEX SETS (OXIRIS) | DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM | KDI | VANTIVE US HEALTHCARE LLC | NA | 24F0086 | 07332414124090 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |