FDA Adverse Event Malfunction Summary report: N

DAVINCI SURGICAL SYSTEM

MDR report key: 2250774 · Received September 7, 2011

Report

Report Number
2250774
Event Type
Malfunction
Date Received
September 7, 2011
Date of Event
February 3, 2011
Report Date
September 7, 2011
Manufacturer
INTUITIVE SURGICAL, INC.
Product Code
NAY
Product Problem
Yes
Report Source
User Facility report
Reporter Location
FL, US

Narratives

Description of Event or Problem · 1

ROBOT FAILURE AND MALFUNCTION DURING ROBOTIC ASSISTED RADICAL PROSTATECTOMY. INSTRUMENTS ATTACHED TO ROBOT ARM, SURGEON AT CONSOLE WITH NO DIFFICULTIES NOTED. ROBOT DOCKED FOR SURGERY AND DISSECTION BEGUN BY SURGEON. DISSECTION ACCOMPLISHED THROUGH PERIVESICAL SPACE WHEN SYSTEM MALFUNCTIONED FOR APPROXIMATELY 30 MINUTES. TWICE, ROBOT FAILED DURING ADDITIONAL DISSECTION REQUIRING REDOCKING OF ROBOT, TURNING OFF UNIT, AND RESTARTING. SYSTEM FINALLY DID NOT REBOOT AND SURGEON HAD TO SWITCH TO OPEN PROCEDURE.======================MANUFACTURER RESPONSE FOR DAVINCI ROBOT, INTUITIVE DAVINCI (PER SITE REPORTER)======================ROBOT NOTED TO HAVE DEFECTIVE PARTS. PARTS REPLACED INCLUDE:-INSTALLED: PART 650340-13-03 (NO OLD PART REMOVED).-REMOVED: 370967-01 (ASSEMPLY, UIP II MEMBRANE & BACKER, ENGLISH). -REMOVED: 380099-01 (ASSEMBLY, UIP, REMOVABLE). NEW PART INSTALLED.FOLLOWING THIS ACTION, MANUFACTURER NOTED NO OTHER ISSUES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 DAVINCI SURGICAL SYSTEM DAVINCI SURGICAL SYSTEM NAY INTUITIVE SURGICAL, INC. IS1000 NONE

Patients

Seq Age Sex Outcome Treatment
1 65 YR OTHER