FDA Adverse Event Injury Summary report: N

SYNCHROMED II

MDR report key: 2250728 · Received September 12, 2011

Report

Report Number
3004209178-2011-07697
Event Type
Injury
Date Received
September 12, 2011
Report Date
September 6, 2011
Manufacturer
MDT PUERTO RICO OPERATIONS CO., JUNCOS
Product Code
LKK
PMA / PMN Number
P860004
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

(B)(4).

Description of Event or Problem · 1

IT WAS INITIALLY REPORTED THAT THE PT HEARD AN ALARM, BUT WAS NOT CONFIRMED BY TELEMETRY. PT HAD RELOCATED AND WAS SEEING A NEW PHYSICIAN. THEY WERE THINKING THE PUMP WAS EMPTY. PT WAS TO GET A REFILL ON (B)(6) 2011 AND ALSO DO A DYE STUDY. IT WAS LATER REPORTED THAT THE PT'S PUMP WAS EMPTY AS THE PT HAD MISSED THE REFILL DATE AND DID NOT SEEK FURTHER ASSISTANCE WITH HER PUMP UNTIL AFTER RELOCATING IN A NEW STATE. THE PT INDICATED THAT SHE HAD EXPERIENCED INCREASED PAIN SYMPTOMS AND HAD BEEN EXPERIENCING THOSE SYMPTOMS SINCE HER PUMP HAD RUN DRY A COUPLE OF MONTHS PRIOR TO RELOCATION. THE PHYSICIAN OPTED TO REPLACE THE ENTIRE SYSTEM. THE NEW PUMP WAS FILLED WITH MORPHINE 5.0 MG/ML AND INITIALLY STARTED AT 0.2599/DAY TO START SINCE PT HAD BEEN WITHOUT DRUG FOR A COUPLE OF MONTHS. MORPHINE WAS ALSO THE DRUG INFUSED BY THE OLD PUMP. A DYE STUDY HAD BEEN CARRIED OUT PRIOR TO REPLACEMENT. THE PHYSICIAN ALSO FOUND THAT THE CATHETER COILED UNDERNEATH PUMP "INDICATING THAT THE PT MAY HAVE BEEN WITHOUT DRUG FOR LONGER THAN WHAT MAY HAVE BEEN EXPECTED BY SIMPLY A MISSED FILL AND SUBSEQUENT EMPTY PUMP". THE EVENT LOGS OF THE OLD PUMP INDICATED THAT A LOW RESERVOIR ALARM AND AN EMPTY RESERVOIR ALARM HAD OCCURRED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 SYNCHROMED II LKK MDT PUERTO RICO OPERATIONS CO., JUNCOS 863740 NA

Patients

Seq Age Sex Outcome Treatment
1 Required Intervention EXPLANTED:| IMPLANTED:| CATHETER: MODEL 8709, LOT #N100896035