RESTORE ULTRA
Report
- Report Number
- 3004209178-2011-07621
- Event Type
- Injury
- Date Received
- September 9, 2011
- Date of Event
- August 18, 2011
- Report Date
- November 10, 2011
- Manufacturer
- MDT PUERTO RICO OPERATIONS CO
- Product Code
- LGW
- PMA / PMN Number
- P840001
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- AR, US
- Reporter Occupation
- PHYSICIAN
Narratives
FINAL DEVICE ANALYSIS OF IMPLANTABLE NEUROSTIMULATOR (INS) REVEALED NO ANOMALY FOUND - INS HAS REDUCED BATTERY CAPACITY DUE TO OVERDISCHARGE. TELEMETRY WAS RESTORED WITH THE DEVICE AFTER ONE PHYSICIAN MODE RECHARGE. FINAL ANALYSIS OF THE EXTENSIONS REVEALED NO SIGNIFICANT ANOMALIES, EXTENSIONS WERE CUT THROUGH, SUSPECTED EXPLANT DAMAGE.
(B)(4). ANALYSIS RESULTS WERE NOT AVAILABLE AT THE TIME OF THIS REPORT. A F/U REPORT WILL BE SENT WHEN ANALYSIS IS COMPLETED.
ADDITIONAL INFORMATION RECEIVED CONFIRMED THE DEVICE HAD BEEN OVERDISCHARGED, AND REPORTED THIS WAS THE SECOND TIME THE DEVICE HAD B EEN IN AN OVERDISCHARGED STATE. A PHYSICIAN MODE RECHARGE WAS ATTEMPTED BUT WAS UNSUCCESSFUL IN RESETTING THE DEVICE.
IT WAS REPORTED THAT THE PT HAD PROGRESSIVE DIFFICULTY WITH RECHARGING THE DEVICE. THE DEVICE WAS IN AN OVER-DISCHARGED STATE AND A TRICKLE CHARGE COULD NOT BE OBTAINED. THE DEVICE WAS EXPLANTED AND THE PT RECOVERED WITHOUT SEQUELA. ADDITIONAL INFO HAS BEEN REQUESTED, BUT WAS NOT AVAILABLE AS OF THE DATE OF THIS REPORT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | RESTORE ULTRA | STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF | LGW | MDT PUERTO RICO OPERATIONS CO | 37712 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 54 YR | Required Intervention | IMPLANTED:| IMPLANTED:| IMPLANTED:| EXPLANTED:| EXPLANTED:| EXPLANTED:| IMPLANTED:| EXPLANTED:| LEAD MODEL 3998 LOT# V131835 IMPLANTED: (B)(4)| EXTENSION: MODEL 37083, LOT# NKC017636N| LEAD: MODEL 3998, LOT# V131835| ACCESSORY: MODEL 37752, LOT# NKA129027N| EXPLANTED:| IMPLANTED:| PROGRAMMER: MODEL 37743, LOT# NKE100343N| EXTENSION: MODEL 37083, LOT# NKC017634N |