FDA Adverse Event Malfunction Summary report: N

UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER

MDR report key: 2250566 · Received September 16, 2011

Report

Report Number
2050012-2011-05570
Event Type
Malfunction
Date Received
September 16, 2011
Date of Event
August 16, 2011
Report Date
August 16, 2011
Manufacturer
BECKMAN COULTER INC.
Product Code
JJE
PMA / PMN Number
K042291
Removal / Correction Number
N/A
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PA, US
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 AND FOUND THE FITTING FOR TUBING 213 ON THE FLUID DISTRIBUTION MANIFOLD WAS LEAKING. THE FSE REPLACED THE FITTING. NO FURTHER LEAKS WERE OBSERVED IN THE INSTRUMENT. REPAIR WAS VERIFIED PER ESTABLISHED PROCEDURES PRIOR TO RETURNING IT INTO SERVICE. (B)(4).

Description of Event or Problem · 1

A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REPORTING A LEAK THAT WAS COMING FROM THE UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER. NO INJURY OR OPERATOR EXPOSURE WAS REPORTED IN THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE JJE BECKMAN COULTER INC. DXC 600 PRO N/A

Patients

Seq Age Sex Outcome Treatment
1