UNICEL® DXC 600 PRO SYNCHRON® CHEMISTRY ANALYZER
MDR report key: 2250566
·
Received September 16, 2011
Report
- Report Number
- 2050012-2011-05570
- Event Type
- Malfunction
- Date Received
- September 16, 2011
- Date of Event
- August 16, 2011
- Report Date
- August 16, 2011
- Manufacturer
- BECKMAN COULTER INC.
- Product Code
- JJE
- PMA / PMN Number
- K042291
- Removal / Correction Number
- N/A
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
A FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED ON (B)(4) 2011 AND FOUND THE FITTING FOR TUBING 213 ON THE FLUID DISTRIBUTION MANIFOLD WAS LEAKING. THE FSE REPLACED THE FITTING. NO FURTHER LEAKS WERE OBSERVED IN THE INSTRUMENT. REPAIR WAS VERIFIED PER ESTABLISHED PROCEDURES PRIOR TO RETURNING IT INTO SERVICE. (B)(4).
Description of Event or Problem · 1
A CUSTOMER CONTACTED BECKMAN COULTER INC. (BEC) REPORTING A LEAK THAT WAS COMING FROM THE UNICEL DXC 600 PRO SYNCHRON CHEMISTRY ANALYZER. NO INJURY OR OPERATOR EXPOSURE WAS REPORTED IN THIS EVENT.