FDA Adverse Event
Malfunction
Summary report: N
MANUAL ORTHOPEDIC SURGICAL INSTRUMENT
MDR report key: 2250522
·
Received September 16, 2011
Report
- Report Number
- 1030489-2011-01172
- Event Type
- Malfunction
- Date Received
- September 16, 2011
- Date of Event
- June 24, 2011
- Report Date
- August 17, 2011
- Manufacturer
- WARSAW ORTHOPEDIC, INC.
- Product Code
- HXX
- PMA / PMN Number
- EXEMPT
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- PHYSICIAN
Narratives
Additional Manufacturer Narrative · 1
(B)(4): THE DEVICE WAS RETURNED FOR EVALUATION. VISUALLY CONFIRMED APPROXIMATELY ~3MM OF THE INSTRUMENT TIP HAS BEEN BROKEN OFF, CONSISTENT WITH INTERFACE DURING USAGE. FRACTURE SURFACE ANALYSIS REVEALS A FAIRLY FLAT FRACTURE SURFACE AND CIRCULAR MATERIAL FLOW, CONSISTENT WITH TORSIONAL OVERLOAD. A REVIEW OF THE DEVICE HISTORY RECORDS DID NOT IDENTIFY ANY APPLICABLE NONCONFORMANCE TO SPECIFICATION.
Description of Event or Problem · 1
IT WAS REPORTED THAT THE THREAD ON THE DRIVER IS WORN OUT. NO PATIENT COMPLICATIONS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | MANUAL ORTHOPEDIC SURGICAL INSTRUMENT | SCREWDRIVER | HXX | WARSAW ORTHOPEDIC, INC. | NA | BM04J010 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|