FDA Adverse Event Injury Summary report: N

NAVI-STAR¿ THERMO-COOL¿ ELECTROPHYSIOLOGY CATHETER

MDR report key: 2250447 · Received September 15, 2011

Report

Report Number
9673241-2011-00095
Event Type
Injury
Date Received
September 15, 2011
Date of Event
August 25, 2011
Report Date
August 29, 2011
Manufacturer
BIOSENSE WEBSTER, INC. (JUAREZ)
Product Code
OAD
PMA / PMN Number
P030031
Removal / Correction Number
NA
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE CATHETER WAS TESTED AND PASSED ELECTRICAL TEST, TEMPERATURE BATH TEST, GENERATOR TEST, PATENCY AND FLOW RATE TEST, DEFLECTION TEST AND FOR VISUAL CHARACTERISTICS INSPECTION. THE ROOT CAUSE OF THE REPORTED EVENT COULD NOT BE DETERMINED. HOWEVER, IT DOES NOT APPEAR TO BE MANUFACTURING RELATED AS THE DEVICE HISTORY RECORD (DHR) WAS REVIEWED AND NO ANOMALIES WERE FOUND RELATED TO THIS COMPLAINT. IN ADDITION, DHR REVIEW VERIFIES THAT THE DEVICE WAS MANUFACTURED IN ACCORDANCE WITH DOCUMENTED SPECIFICATION AND PROCEDURES. THE COMPLAINT CANNOT BE CONFIRMED.

Additional Manufacturer Narrative · 1

THE EVENT WAS LIFE THREATENING AND THE DAMAGE TO HEART WERE MODERATE. PHYSICIAN/CUSTOMER'S OPINION REGARDING THE CAUSALITY OF ADVERSE EVENT WAS PROCEDURE-RELATED. THE CONCOMITANT DEVICES: CARTO 3 SYSTEM, US CATALOG #: FG540000, (B)(4); STOCKERT 70 SYSTEM, US CATALOG #: S7001, (B)(4); COOL FLOW PUMP, U.S. SHIP KIT, US CATALOG #: CFP002, (B)(4). (B)(4)

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LEFT SIDED ATRIAL TACHYCARDIA, THE PATIENT EXPERIENCED A TAMPONADE. THE PATIENT'S BLOOD PRESSURE DECREASED AND A PERICARDIOCENTESIS WAS PERFORMED EXTRACTING 300 CC OF FLUID. THE PATIENT WAS GIVEN UNKNOWN MEDICATIONS, THEN STABILIZED AND RETURNED TO THE HOLDING ROOM. THE DRAIN WAS LEFT IN OVERNIGHT AND THE PATIENT WAS DISCHARGED THREE DAYS LATER. THE FOLLOWING BWI PRODUCTS WERE INVOLVED CARTO 3, STOCKERT, COOL FLOW PUMP, NAVISTAR CATHETER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 NAVI-STAR¿ THERMO-COOL¿ ELECTROPHYSIOLOGY CATHETER CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FLUTTER OAD BIOSENSE WEBSTER, INC. (JUAREZ) D-1197-17-S 15393107M

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R