MINIMED QUICK-SET
Report
- Report Number
- 3003442380-2025-11553
- Event Type
- Injury
- Date Received
- July 15, 2025
- Date of Event
- June 15, 2025
- Report Date
- July 11, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244017573
- PMA / PMN Number
- K160648
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
E1: PATIENT CITY: (B)(6). PATIENT COUNTRY: THE UNITED STATES.
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: THE INFORMATION IN THIS COMPLAINT (B)(4) HAS BEEN EVALUATED FOR THE MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION. THE BATCH 6010426 IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. COMPLAINT INVESTIGATION: THE REFERENCE SAMPLES CANNOT BE TESTED BECAUSE THERE WAS NO SPECIFIC MALFUNCTION TO INVESTIGATE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010426 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 82 PACKAGING IN THE MULTIVAC 12, ON 30/NOV/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WERE IDENTIFIED RELATED TO THE MALFUNCTION REPORTED, NO MAINTENANCE EVENTS WERE RECORDED. TRENDING: A QUERY WAS RUN IN DATABASE ON 11/JUL/2025 AGAINST MALFUNCTION CODE NO MALFUNCTION BASED ON COMPLAINT INFORMATION, HARM CODE UNTREATED DIABETIC KETOACIDOSIS OR ISOLATED ELEVATED BLOOD GLUCOSE WHICH THE PATIENT IS UNABLE TO SELF-MANAGE REQUIRING ADVANCED LIFE SUPPORT TO PREVENT PERMANENT ORGAN DAMAGE (ELEVATED BLOOD GLUCOSE LEVEL, PRESENCE OF KETONES AND SYMPTOMS E.G., NAUSEA, VOMITING, ABDOMINAL PAIN, CONFUSION) AND LOT 6010426 AND ONE OTHER COMPLAINTS HAVE BEEN REGISTERED IN DATABASE FOR THE SAME LOT NUMBER, HARM CODE AND MALFUNCTION CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION, ONE OTHER COMPLAINT RECEIVED ON THE LOT IN QUESTION, HARM CODE AND MALFUNCTION CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION OR CAPA PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT WAS HOSPITALIZED (B)(6) 2025 DUE TO HYPERGLYCEMIA. THE BLOOD GLUCOSE LEVEL WAS 475 MG/DL AT THE TIME OF EVENT AND THE PATIENT GOT TREATED WITH INTRAVENOUS DRIP. LENGTH OF HOSPITALIZATION WAS FOR LESS THAN 24 HOURS. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 137661 | MINIMED QUICK-SET | UNO QUICK-SET 60/9 SC1 MECA | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | MMT-397A | 6010426 | 05705244017573 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown | Required Intervention |