COLLEAGUE
Report
- Report Number
- 6000001-2011-24974
- Date Received
- September 15, 2011
- Date of Event
- August 27, 2011
- Report Date
- August 30, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Removal / Correction Number
- 6000001-12/1/08-002-C
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- FL, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM. THE DEVICE IS CURRENTLY IN THE PROCESS OF BEING EVALUATED ON SITE BY A BAXTER FIELD SERVICE TECHNICIAN. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE.
(B)(4). EVALUATION SUMMARY: THE REPORTED CONDITION OF A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY WAS CONFIRMED DURING PRODUCT EVALUATION. THE ROOT CAUSE OF THIS CONDITION WAS ASSIGNED TO DEPLETED MAIN BATTERIES RESULTING FROM USER ERROR. THE MAIN BATTERIES AND BATTERY HARNESS WERE REPLACED TO CORRECT THIS CONDITION. ADDITIONAL INFORMATION: THIS ISSUE IS BEING INVESTIGATED THROUGH CAPA.
THE FACILITY REPRESENTATIVE CONTACTED A TECHNICAL SERVICE REPRESENTATIVE TO REPORT A COLLEAGUE INFUSION PUMP WITH A DAMAGED BATTERY; THIS CONDITION HAD THE POTENTIAL TO INTERRUPT DELIVERY. IT IS UNKNOWN WHEN THIS EVENT OCCURRED. THERE WERE NO REPORTS OF PATIENT INVOLVEMENT, PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE EVENT IN ASSOCIATION WITH THIS EVENT. THIS INVOLVED A REMEDIATED COLLEAGUE INFUSION PUMP WITH USER INTERFACE MODULE MASTER SOFTWARE VERSION 6.13.90. NO ADDITIONAL INFORMATION IS AVAILABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | COLLEAGUE | PUMP, INFUSION | FRN | BAXTER HEALTHCARE - SINGAPORE |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |