SYNCHRON® LX20 PRO CLINICAL SYSTEM
Report
- Report Number
- 2050012-2011-05280
- Event Type
- Malfunction
- Date Received
- September 15, 2011
- Date of Event
- August 18, 2011
- Report Date
- August 18, 2011
- Manufacturer
- BECKMAN COULTER, INC.
- Product Code
- JJE
- PMA / PMN Number
- K011213
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NJ, US
- Reporter Occupation
- OTHER
Narratives
QC HAD BEEN WITHIN THE ESTABLISHED RANGES. THE INSTRUMENT WAS SHOWING REAGENT PROBE B LEVEL SENSE ERRORS AND "COMPARTMENT C NOT EMPTY" ON TRIGLYCERIDE TEST ALTHOUGH THE CARTRIDGE WAS PROPERLY PREPARED AND THE COMPARTMENT C WAS EMPTY. THE CUSTOMER FOUND LIQUID UNDER THE TRAY OVER THE REAGENT CAROUSEL AREA, AND BOTTOM OF THE COLLAR WASH FOR REAGENT PROBE B. ON (B)(4) 2011, BEC FIELD SERVICE ENGINEER (FSE) PERFORMED SERVICE ON THE INSTRUMENT. THE FSE REPLACED THE LEVEL SENSE BEAD AND PERFORMED ALL NECESSARY ALIGNMENTS. THE FSE VERIFIED REPAIR PER ESTABLISHED PROCEDURES. (B)(4).
A CUSTOMER REPORTED TO BECKMAN COULTER, INC. (BEC) THAT ERROR MESSAGES AND FLUID LEAKS WERE OBSERVED ON SYNCHRON LX20 PRO CLINICAL SYSTEM. BEC CUSTOMER TECHNICAL SUPPORT (CTS) ASKED THE CUSTOMER TO DISABLE THE CARTRIDGE CHEMISTRY SIDE OF THE INSTRUMENT. THE CUSTOMER INDICATED THAT NO ERRONEOUS RESULTS WERE GENERATED AND PATIENT TREATMENT WAS NOT AFFECTED. MSDS WAS NOT REVIEWED, BUT THE FACILITY HAS AN EXPOSURE CONTROL PLAN. THE OPERATOR DID NOT SEEK MEDICAL ATTENTION AND THERE WAS NO EFFECT TO PATIENT OR USER.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | SYNCHRON® LX20 PRO CLINICAL SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER, INC. | LX20 PRO | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |