FDA Adverse Event Malfunction Summary report: N

BILIARY BALLOON DILATATION CATHETER

MDR report key: 2250209 · Received September 15, 2011

Report

Report Number
3005099803-2011-03203
Event Type
Malfunction
Date Received
September 15, 2011
Report Date
August 19, 2011
Manufacturer
BOSTON SCIENTIFIC - GALWAY
Product Code
FGE
PMA / PMN Number
K910931
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND NO VISIBLE ISSUES WITH THE CATHETER OF THE DEVICE. HOWEVER, THROUGH FUNCTIONAL TESTING A PINHOLE WAS NOTED TO THE BALLOON MATERIAL. A SMALL TEAR WAS ALSO NOTED TO THE BALLOON MATERIAL. THE INVESTIGATION RESULTS ARE CONSISTENT WITH THE EVENT DESCRIPTION. THE REPORTED EVENT OF BALLOON INFLATION DIFFICULTY WAS CONFIRMED, AS A PINHOLE WAS NOTED TO THE BALLOON. THIS FAILURE LIKELY OCCURRED DUE TO ANATOMICAL/PROCEDURAL FACTORS WHICH LIMITED THE PERFORMANCE OF THE BALLOON. THEREFORE, THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) WAS PERFORMED AND CONFIRMED THAT THIS DEVICE MET ALL MATERIAL, ASSEMBLY AND PERFORMANCE SPECIFICATIONS. A SEARCH OF THE COMPLAINT DATABASE REVEALED THAT NO SIMILAR COMPLAINTS EXIST FOR THE SPECIFIED LOT.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A MAXFORCE BALLOON WAS USED DURING AN ENDOSCOPIC RETROGRADE CHOLANGIOPANCREATOGRAPHY (ERCP) IN THE DUODENAL PAPILLA. ACCORDING TO THE COMPLAINT, DURING THE PROCEDURE, THE BALLOON ONLY INFLATED A LITTLE. IT WAS REPORTED THAT THE BALLOON DID NOT BURST. THROUGH THE INVESTIGATION RESULTS, IT WAS REPORTED THAT THE BALLOON HAD A PINHOLE. THE PROCEDURE WAS COMPLETED WITH ANOTHER MAXFORCE BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THE EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 BILIARY BALLOON DILATATION CATHETER CATHETER, BILIARY, DIAGNOSTIC FGE BOSTON SCIENTIFIC - GALWAY M00567420 13875388

Patients

Seq Age Sex Outcome Treatment
1