FDA Adverse Event Malfunction Summary report: N

RAPID EXCHANGE BILIARY BALLOON DILATATION CATHETER

MDR report key: 2250159 · Received September 15, 2011

Report

Report Number
3005099803-2011-03195
Event Type
Malfunction
Date Received
September 15, 2011
Date of Event
July 13, 2011
Report Date
August 19, 2011
Manufacturer
BOSTON SCIENTIFIC - CORK
Product Code
FGE
PMA / PMN Number
K001338
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4): VISUAL EXAMINATION OF THE RETURNED DEVICE FOUND NO DAMAGE TO THE CATHETER OF THE DEVICE. HOWEVER, THROUGH FUNCTIONAL TESTING IT WAS REVEALED THAT A PINHOLE WAS PRESENT IN THE BALLOON MATERIAL. THE INVESTIGATION RESULTS ARE CONSISTENT WITH THE EVENT DESCRIPTION. THE REPORTED DEFECT OF BALLOON PINHOLE WAS CONFIRMED. THIS FAILURE LIKELY OCCURRED DUE TO ANATOMICAL/PROCEDURAL FACTORS WHICH LIMITED THE PERFORMANCE OF THE BALLOON. THEREFORE, THE MOST PROBABLE ROOT CAUSE IS OPERATIONAL CONTEXT. A SEARCH OF THE COMPLAINT DATABASE REVEALED THAT NO SIMILAR COMPLAINTS EXIST FOR THE SPECIFIED LOT.

Description of Event or Problem · 1

IT WAS REPORTED TO BOSTON SCIENTIFIC CORPORATION THAT A RX DILATATION BALLOON WAS USED DURING A BALLOON DILATATION PROCEDURE PERFORMED ON (B)(6), 2011. ACCORDING TO THE COMPLAINT, IT WAS REPORTED THAT THE BALLOON WAS ATTEMPTED TO BE INFLATED, BUT WAS UNSUCCESSFUL. IT WAS CONFIRMED THAT THE BALLOON HAD A LEAK DUE TO A PINHOLE IN THE BALLOON MATERIAL. ATTEMPTS TO OBTAIN ADDITIONAL PATIENT AND PROCEDURE INFORMATION HAVE BEEN UNSUCCESSFUL. IF ANY FURTHER RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL MDR WILL BE FILED. THE PROCEDURE WAS COMPLETED WITH ANOTHER RX DILATATION BALLOON. THERE WERE NO PATIENT COMPLICATIONS REPORTED AS A RESULT OF THE EVENT. THE PATIENT'S CONDITION AT THE CONCLUSION OF THE PROCEDURE WAS REPORTED TO BE GOOD.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 RAPID EXCHANGE BILIARY BALLOON DILATATION CATHETER CATHETER, BILIARY, DIAGNOSTIC FGE BOSTON SCIENTIFIC - CORK M00545940 14304103

Patients

Seq Age Sex Outcome Treatment
1