COLLEAGUE
Report
- Report Number
- 6000001-2011-24918
- Event Type
- Malfunction
- Date Received
- September 15, 2011
- Date of Event
- August 26, 2011
- Report Date
- August 29, 2011
- Manufacturer
- BAXTER HEALTHCARE - SINGAPORE
- Product Code
- FRN
- PMA / PMN Number
- K063696
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- VA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
(B)(4). EVALUATION SUMMARY: THE CONDITION OF A COLLEAGUE INFUSION PUMP THAT EXPERIENCED A DEPLETED BATTERY WAS CONFIRMED AND REPRODUCED BY BAXTER PERSONNEL DURING PRODUCT EVALUATION. THE ASSIGNABLE CAUSE FOR THE REPORTED PROBLEM WAS DETERMINED TO BE DEPLETED BATTERIES RESULTING FROM USER ERROR. THIS CONDITION WAS RESOLVED BY REPLACING THE MAIN BATTERIES AND BATTERY HARNESS.
(B)(4). THE DEVICE WAS RETURNED TO BAXTER AND IS CURRENTLY AWAITING EVALUATION. A FOLLOW-UP REPORT WILL BE FILED UPON COMPLETION OF THE EVALUATION OR IF ANY ADDITIONAL DETAILS BECOME AVAILABLE. SIMILAR REPORTS HAVE BEEN RECEIVED FOR THE REPORTED PROBLEM.
THE FACILITY REPORTED A COLLEAGUE INFUSION PUMP WITH A DEPLETED BATTERY. ACCORDING TO THE FACILITY, THIS EVENT OCCURRED UPON POWER-UP IN THE MEDICAL SURGICAL DEPARTMENT. THIS EVENT MAY HAVE INTERRUPTED DELIVERY. THERE WAS NO PATIENT INVOLVEMENT; THEREFORE, NO PATIENT INJURY, MEDICAL INTERVENTION, OR ADVERSE REACTION IS ASSOCIATED WITH THE REPORTED CONDITION. NO ADDITIONAL INFORMATION IS AVAILABLE. THIS DEVICE IS A REMEDIATED COLLEAGUE PUMP WITH A USER INTERFACE MODULE SOFTWARE VERSION OF 6.13.90.