FDA Adverse Event Injury Summary report: N

PERCLOSE¿ PROSTYLE¿

MDR report key: 22500715 · Received July 14, 2025

Report

Report Number
2024168-2025-07234
Event Type
Injury
Date Received
July 14, 2025
Date of Event
June 24, 2025
Report Date
August 26, 2025
Manufacturer
ABBOTT VASCULAR INC.
Product Code
MGB
UDI-DI
08717648344039
PMA / PMN Number
P960043
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

MANUFACTURER'S INVESTIGATION IS STILL PENDING AT THIS TIME. RESULTS AND CONCLUSIONS WILL BE PROVIDED IN THE FINAL REPORT. THE ADDITIONAL DEVICE REFERENCED IN B5 IS FILED UNDER A SEPARATE MEDWATCH REPORT NUMBER.

Additional Manufacturer Narrative · 0

A VISUAL INSPECTION WAS PERFORMED ON THE RETURNED DEVICE. THE REPORTED NEEDLE TO CUFF MISS WAS CONFIRMED. PRODUCTION RECORD AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) REVIEWS WERE PERFORMED AND REVEALED NO INDICATION OF A PRODUCT QUALITY ISSUE. A REVIEW OF THE PRODUCTION RECORD AND CORRECTIVE AND PREVENTIVE ACTIONS (CAPA) DATABASE FOR THE REPORTED LOT REVEALED NO RELATED MANUFACTURING NONCONFORMITIES. BASED ON THE REPORTED INFORMATION AND THE OBSERVATIONS FROM THE RETURNED ANALYSIS, THE INVESTIGATION DETERMINED THAT THE REPORTED ISSUE AND SUBSEQUENT TREATMENT APPEAR TO BE RELATED TO OPERATIONAL CONTEXT. IT IS LIKELY THAT AN INTERACTION OF THE DEVICE WITH PATIENT ANATOMY OR INABILITY TO MAINTAIN POSITION/STABILITY OF THE DEVICE DURING DEPLOYMENT DUE TO CIRCUMSTANCES OF THE PROCEDURE. BASED ON THE RESULTS OF THE COMPLAINT INVESTIGATION, THERE IS NO INDICATION OF A PRODUCT QUALITY ISSUE WITH RESPECT TO THE DESIGN, MANUFACTURE, OR LABELING OF THE DEVICE. D4: LOT NUMBER UPDATED FROM 5032642 TO 5021641.

Description of Event or Problem · 0

IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF A CALCIFIED COMMON FEMORAL ARTERY USING THE PRE-CLOSE TECHNIQUE VIA A 6F SHEATH HOLE PRIOR TO AN INTERVENTIONAL ABDOMINAL AORTIC ANEURYSM (AAA) PROCEDURE. REPORTEDLY, A CUFF MISS [SUTURE RETRIEVAL ISSUE] OCCURRED WITH TWO PROSTYLE DEVICES. THE SUTURES OF TWO NEW PROSTYLE DEVICES WERE SUCCESSFULLY PRE-PLACED. THE SHEATH WAS UPSIZED TO A 16F, AND THE INTERVENTIONAL PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED WITH THE PRE-PLACED PROSTYLE SUTURES. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Description of Event or Problem · 0

SUBSEQUENT TO THE PREVIOUSLY FILED REPORT, THE FOLLOWING INFORMATION WAS RECEIVED: IT WAS REPORTED THAT THIS WAS AN ARTERIOTOMY CLOSURE OF A CALCIFIED COMMON FEMORAL ARTERY USING THE PRE-CLOSE TECHNIQUE VIA A 6F SHEATH HOLE PRIOR TO AN INTERVENTIONAL ABDOMINAL AORTIC ANEURYSM (AAA) PROCEDURE. REPORTEDLY, A CUFF MISS [SUTURE RETRIEVAL ISSUE] OCCURRED WITH TWO PROSTYLE DEVICES. THE SUTURES OF TWO NEW PROSTYLE DEVICES WERE SUCCESSFULLY PRE-PLACED. THE SHEATH WAS UPSIZED TO A 16F, AND THE INTERVENTIONAL PROCEDURE WAS COMPLETED. HEMOSTASIS WAS ACHIEVED WITH THE PRE-PLACED PROSTYLE SUTURES. THERE WAS NO REPORTED ADVERSE PATIENT SEQUELA AND NO REPORTED CLINICALLY SIGNIFICANT DELAY IN THE PROCEDURE OR THERAPY. NO ADDITIONAL INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
400896 PERCLOSE¿ PROSTYLE¿ DEVICE, HEMOSTASIS, VASCULAR MGB ABBOTT VASCULAR INC. 12773-02 5021641 08717648344039

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown Required Intervention