FDA Adverse Event
Injury
Summary report: N
EGIA ULTRA UNIVERSAL SHORT STAPLER
MDR report key: 2250059
·
Received September 9, 2011
Report
- Report Number
- 1219930-2011-00773
- Event Type
- Injury
- Date Received
- September 9, 2011
- Date of Event
- August 25, 2011
- Report Date
- August 26, 2011
- Manufacturer
- UNITED STATES SURGICAL
- Product Code
- GDW
- PMA / PMN Number
- K083519
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
(B)(4).
Description of Event or Problem · 1
PROCEDURE TYPE: PROXIMAL PANCREATICODUODENECTOMY. ACCORDING TO THE REPORTER: WHEN APPROACHED ON DUODENUM, THE JAWS WOULD NOT CLOSE TIGHT ENOUGH. IT WAS FORCED CLOSED AND FIRING WAS DONE, BUT A BROKEN SOUND WAS HEARD. MOREOVER, FIRING WAS NOT COMPLETED. ADDITIONAL RESECTION OF TISSUE WAS NEEDED. ANOTHER DEVICE WAS USED TO COMPLETE THE PROCEDURE. NO BLEEDING OCCURRED. NOTHING FELL INTO THE PATIENT CAVITY. OPERATIVE TIME WAS NOT EXTENDED MORE THAN 30 MINUTES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | EGIA ULTRA UNIVERSAL SHORT STAPLER | DISPOSABLE SURGICAL STAPLING DEVICE | GDW | UNITED STATES SURGICAL | N1E0811UKX |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Disability | EGIA 60 ARTICULATING MED/THICK SULU:| EXP: 05/31/2016, MANUF: 05/2011, (B)(4)| CATALOG #: EGIA60AMT, LOT# N1E0481UL |