FDA Adverse Event Injury Summary report: N

INRATIO

MDR report key: 2248366 · Received September 8, 2011

Report

Report Number
2027969-2011-01933
Event Type
Injury
Date Received
September 8, 2011
Date of Event
August 8, 2011
Report Date
September 8, 2011
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
GJS
PMA / PMN Number
K021923
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

PT WAS ON LOVENOX. LOVENOX IS A CLASS OF ANTITHROMBOTIC AGENTS KNOWN AS LOW-MOLECULAR-WEIGHT HEPARINS (LMWH). PER PRODUCT USER GUIDE - LIMITATIONS, "THIS TEST SHOULD NOT BE USED FOR PTS ON HEPARIN THERAPY." PT IS ALSO ON SEVERAL OTHER MEDICATIONS. PER PRODUCT USER GUIDE, "CERTAIN PRESCRIPTION DRUGS AND OVER-THE-COUNTER MEDICATIONS (E.G., ANTIBIOTICS, PAIN RELIEVERS) CAN AFFECT THE ACTION OF ORAL ANTICOAGULANTS. STARTING, STOPPING OR CHANGING THE DOSE CAN AFFECT THE INR VALUE." CUSTOMER SOMETIMES MILKS FINGER. PER PRODUCT USER GUIDE - PRECAUTIONS AND WARNINGS, "DO NOT USE REPETITIVE PRESSURE TO COLLECT THE SAMPLE. "SQUEEZING THE FINGER STICK SITE EXCESSIVELY (MILKING) RELEASES INTERSTITIAL FLUID AND MAY CAUSE IN ACCURATE RESULTS. QC ERRORS ARE DESIGNED TO BE GENERATED IF THE STRIP HAS BEEN EXPOSED TO ADVERSE ENVIRONMENTAL CONDITION. THERE IS NO INFO IN THE COMPLAINT TO INDICATE STRIP EXPOSURE. THESE ERRORS ALSO ARISE IF THERE IS A SAMPLING OR TECHNIQUE PROBLEM. THIS FAILURE MODE WILL CONTINUE TO BE MONITORED TO DETERMINE IF CORRECTIVE ACTION IS WARRANTED. NES ERROR OCCURS WHEN THERE IS NOT ENOUGH SAMPLE APPLIED TO THE TEST STRIPS TO FILL THE TEST AREA, AND/OR STRIP CHANNEL IS BLOCKED. ONE HUNDRED THIRTEEN TOTAL COMPLAINTS HAVE BEEN REPORTED FOR LOT #253024 YIELDING A COMPLAINT RATE OF 0.188%. THIS REACHES THE ACTION THRESHOLD OF 0.1%. BRICK LOT NUMBER #246550 HAS STRIP CODE TS4XB; MATERIAL WAS SPLIT INTO THREE DIFFERENT LOTS: LOT #253024 INTO 12 PACKS (SMALLER LOT). LOT #243394 INTO 12 PACKS (SMALLER LOT), LOT #253028 INTO 48 PACKS (LARGER LOT). THEREFORE, 217 TOTAL COMPLAINTS HAVE BEEN REPORTED FOR LOT NUMBER'S 253024, 243394, 253028 YIELDING A COMPLAINT RATE OF 0.065%. DUE TO THIS LOW OCCURRENCE RATE, BELOW THE ACTION THRESHOLD OF 0.1%, CORRECTIVE ACTION IS NOT REQUIRED AT THIS TIME.

Description of Event or Problem · 1

CALLER ALLEGED DISCREPANT RESULTS USING THE INRATIO METER: RESULTS AS FOLLOWS: DATE; (B)(6) 2011, INRATIO: QC1+2; RE-TEST: QC1+2; RE-TEST: NES. PT SELF TESTER WAS UNABLE TO GET A RESULT, SO NURSE DID A VENOUS DRAW AND RESULTS CAME BACK AT 18.X. PT WAS IMMEDIATELY ADMITTED TO THE HOSPITAL FROM (B)(6) 2011-(B)(6) 2011. PT'S SON STATES THAT THERE WAS A SMALL BRUISE ON HER LEG BECAUSE SHE HAD PUMPED INTO SOMETHING. PT HAD FALLEN AND BROKE HER BACK SO SHE WAS IN A REHAB/NURSING HOME FROM (B)(6) 2011-(B)(6) 2011. WHILE IN THE NURSING HOME, SHE WAS GIVEN LONENOX INJECTIONS. SON DOES NOT REMEMBER WHEN THE INJECTIONS STROPPED, HOWEVER, HE KNOWS THAT SHE WAS NO LONGER TAKING LOVENOX AFTER LEAVING THE NURSING HOME. PT SOMETIME MILKS THE FINGER.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST GJS ALERE SAN DIEGO, INC. 100071 253024

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization