LIBERTY CYCLER
Report
- Report Number
- 2937457-2011-00014
- Event Type
- Malfunction
- Date Received
- August 23, 2011
- Date of Event
- July 25, 2011
- Report Date
- July 25, 2011
- Manufacturer
- FRESENIUS MEDICAL CARE NORTH AMERICA
- Product Code
- FKX
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- UNKNOWN
Narratives
CYCLER PASSED THE SCALE CALIBRATION CHECK AND A VALVE ACTUATION TEST. THE WEIGHTED AND DISPLAYED FILL VOLUMES WERE WITHIN THE 2% TOLERANCE. THE REPORTED DC (DRAIN COMPLICATION), FC (FILL COMPLICATION) OR M65 (SCALE READING ERROR) WERE NOT DUPLICATED DURING SIMULATED TREATMENT. (B)(4). THE STN # K043363.
PATIENT IS ON CCPD AND PATIENT'S WIFE REPORTED THAT THEY WERE UP ALL NIGHT WITH ALARMS. SHE SAID THAT THE CYCLER SHOWED THAT PATIENT DID NOT FILL ENOUGH BUT HE FELT FULL SO HE BYPASSED THE FILLS. TX SETTINGS ARE CYCLE BASED CCPD, 4 FILLS OF 2,000 ML., LAST FILL 0. TX DETAILS: FILL = 675, DRAIN = 3334; FILL = 1212, DRAIN = 4793; FILL = 683, DRAIN = 1055; FILL = 594, DRAIN = 4026. THE REMAINING SOLUTION FROM THE TWO 5-L BAGS USED WAS 3700 ML. INDICATING THAT TOTAL SOLUTION USED WAS ABOUT 6,300 ML. TOTAL FILL VOLUME RECORDED WAS 3164. TOTAL DRAIN VOLUME RECORDED WAS 13208.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | LIBERTY CYCLER | PERITONEAL DIALYSIS CYCLER | FKX | FRESENIUS MEDICAL CARE NORTH AMERICA | LIB | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 42 YR | CASSETTE SET| PD SOLUTION |