FDA Adverse Event Malfunction Summary report: N

LIBERTY CYCLER

MDR report key: 2248200 · Received August 23, 2011

Report

Report Number
2937457-2011-00014
Event Type
Malfunction
Date Received
August 23, 2011
Date of Event
July 25, 2011
Report Date
July 25, 2011
Manufacturer
FRESENIUS MEDICAL CARE NORTH AMERICA
Product Code
FKX
PMA / PMN Number
NA
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

CYCLER PASSED THE SCALE CALIBRATION CHECK AND A VALVE ACTUATION TEST. THE WEIGHTED AND DISPLAYED FILL VOLUMES WERE WITHIN THE 2% TOLERANCE. THE REPORTED DC (DRAIN COMPLICATION), FC (FILL COMPLICATION) OR M65 (SCALE READING ERROR) WERE NOT DUPLICATED DURING SIMULATED TREATMENT. (B)(4). THE STN # K043363.

Description of Event or Problem · 1

PATIENT IS ON CCPD AND PATIENT'S WIFE REPORTED THAT THEY WERE UP ALL NIGHT WITH ALARMS. SHE SAID THAT THE CYCLER SHOWED THAT PATIENT DID NOT FILL ENOUGH BUT HE FELT FULL SO HE BYPASSED THE FILLS. TX SETTINGS ARE CYCLE BASED CCPD, 4 FILLS OF 2,000 ML., LAST FILL 0. TX DETAILS: FILL = 675, DRAIN = 3334; FILL = 1212, DRAIN = 4793; FILL = 683, DRAIN = 1055; FILL = 594, DRAIN = 4026. THE REMAINING SOLUTION FROM THE TWO 5-L BAGS USED WAS 3700 ML. INDICATING THAT TOTAL SOLUTION USED WAS ABOUT 6,300 ML. TOTAL FILL VOLUME RECORDED WAS 3164. TOTAL DRAIN VOLUME RECORDED WAS 13208.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIBERTY CYCLER PERITONEAL DIALYSIS CYCLER FKX FRESENIUS MEDICAL CARE NORTH AMERICA LIB NA

Patients

Seq Age Sex Outcome Treatment
1 42 YR CASSETTE SET| PD SOLUTION