FDA Adverse Event Malfunction Summary report: N

KYPHON XPANDER¿ INFLATABLE BONE TAMP

MDR report key: 22477664 · Received July 11, 2025

Report

Report Number
9617601-2025-00613
Event Type
Malfunction
Date Received
July 11, 2025
Date of Event
June 19, 2025
Report Date
September 9, 2025
Manufacturer
MEDTRONIC MEXICO S. DE R.L. DE CV
Product Code
HRX
UDI-DI
00763000311919
PMA / PMN Number
K041454
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

G2: COUNTRY OF ORIGIN IS JAPAN. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Additional Manufacturer Narrative · 0

H3: VISUAL AND OPTICAL INSPECTION REVEALED THE TIP OF THE BALLOON HAS BEEN DAMAGED. THE UPPER PORTION OF THE BALLOON HAS BEEN SLICED. THIS DAMAGE IS CONSISTENT WITH THE BALLOON COMING IN CONTACT WITH BONE SPURS WHEN THE BALLOON IS INFLATED IN THE VERTEBRAL BODY. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.

Description of Event or Problem · 0

INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING L3 BALLOON KYP HOPLASTY (BKP) FOR PRIMARY OSTEOPOROSIS (COMPRESSION FRACTURE). IT WAS REPORTED THAT THE BALLOON RUPTURED WITHOUT TRIGGER DURING BALLOON INFLATION. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
894013 KYPHON XPANDER¿ INFLATABLE BONE TAMP ARTHROSCOPE HRX MEDTRONIC MEXICO S. DE R.L. DE CV K09A UNKNOWN 00763000311919

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown