KYPHON XPANDER¿ INFLATABLE BONE TAMP
Report
- Report Number
- 9617601-2025-00613
- Event Type
- Malfunction
- Date Received
- July 11, 2025
- Date of Event
- June 19, 2025
- Report Date
- September 9, 2025
- Manufacturer
- MEDTRONIC MEXICO S. DE R.L. DE CV
- Product Code
- HRX
- UDI-DI
- 00763000311919
- PMA / PMN Number
- K041454
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JA
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
G2: COUNTRY OF ORIGIN IS JAPAN. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
H3: VISUAL AND OPTICAL INSPECTION REVEALED THE TIP OF THE BALLOON HAS BEEN DAMAGED. THE UPPER PORTION OF THE BALLOON HAS BEEN SLICED. THIS DAMAGE IS CONSISTENT WITH THE BALLOON COMING IN CONTACT WITH BONE SPURS WHEN THE BALLOON IS INFLATED IN THE VERTEBRAL BODY. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
INFORMATION WAS RECEIVED FROM HEALTHCARE PROVIDER (HCP) VIA A MANUFACTURER REPRESENTATIVE REGARDING A PATIENT HAVING L3 BALLOON KYP HOPLASTY (BKP) FOR PRIMARY OSTEOPOROSIS (COMPRESSION FRACTURE). IT WAS REPORTED THAT THE BALLOON RUPTURED WITHOUT TRIGGER DURING BALLOON INFLATION. THERE WERE NO PATIENT SYMPTOMS REPORTED. THERE WERE NO FURTHER COMPLICATIONS REPORTED REGARDING THE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 894013 | KYPHON XPANDER¿ INFLATABLE BONE TAMP | ARTHROSCOPE | HRX | MEDTRONIC MEXICO S. DE R.L. DE CV | K09A | UNKNOWN | 00763000311919 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |