FDA Adverse Event Injury Summary report: N

INTELECT XT 2CH COMBO PKG

MDR report key: 2247676 · Received September 8, 2011

Report

Report Number
9616086-2011-00099
Event Type
Injury
Date Received
September 8, 2011
Date of Event
August 10, 2011
Report Date
September 9, 2011
Manufacturer
DJO GLOBAL
Product Code
GZJ
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

UNIT BURNED PATIENTS. THE DEVICE EVALUATION REPORT INDICATED THAT NO DAMAGE WAS FOUND THE UNIT AND LEAD CHANNELS ARE IN LIKE NEW CONDITION. ESTIM TREATMENTS ARE EXACT REPLICATIONS OF THE DESIRED SIGNALS. THE LEAD WIRES IMPEDANCE MEASURED TO SPECIFICATIONS, CONNECTIONS WERE NOT INTERMITTENT OR OPEN. POWER LEVEL OUTPUT WAS WITHIN TOLERANCES. NO ERRORS OCCURRED DURING TREATMENT. HI-POT AND LEAKAGE CURRENT PASS ELECTRICAL SAFETY TESTING. LEAKAGE CURRENT AND GROUND CONTINUITY HAVE MET SPECIFICATION REQUIREMENTS. THE UNIT PASSED INSPECTION. THE WAVEFORMS APPEARED AS NORMAL, AND THE UNIT MAINTAINED NORMAL FUNCTIONALITY. NO PROBABLE CAUSE OF AN INCIDENT COULD BE DUPLICATED AND THE UNIT PASSES QC INSPECTION.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INTELECT XT 2CH COMBO PKG ELECTROTHERAPY GZJ DJO GLOBAL

Patients

Seq Age Sex Outcome Treatment
1 Other