FDA Adverse Event Malfunction Summary report: N

GLP CENTRIFUGE MODULE

MDR report key: 22476316 · Received July 10, 2025

Report

Report Number
3023268435-2025-00010
Event Type
Malfunction
Date Received
July 10, 2025
Date of Event
June 30, 2025
Report Date
August 22, 2025
Manufacturer
ABBOTT AUTOMATION SOLUTIONS GMBH
Product Code
JQP
UDI-DI
00380740186647
PMA / PMN Number
K213486
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
UK
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

AN EVALUATION IS IN PROCESS. A FOLLOW-UP REPORT WILL BE SUBMITTED WHEN THE EVALUATION IS COMPLETE. THIS REPORT IS BEING FILED ON AN INTERNATIONAL PRODUCT, LIST NUMBER 06Q03 THAT HAS A SIMILAR PRODUCT DISTRIBUTED IN THE US, LIST NUMBER 04Z96, WITH 510K NUMBER K213486.

Additional Manufacturer Narrative · 0

THE GLP CENTRIFUGE MODULE (CM) SERIAL (B)(6) WAS EVALUATED. PRINTED CIRCUIT BOARD (PCB) ROBOT DISTRIBUTION 5 WAS FOUND TO HAVE ELECTRICAL DAMAGE ISOLATED TO A SINGLE ELECTRICAL COMPONENT. PCB ROBOT DISTRIBUTION 5 WAS REPLACED, WHICH RESOLVED THE CUSTOMER REPORTED EVENT. DATA AND INFORMATION PROVIDED BY THE CUSTOMER WERE REVIEWED AND SUPPORT THE COMPLAINT ISSUE WITHOUT INDICATION FOR ANY ADDITIONAL ISSUE. A REVIEW OF TRACKING AND TRENDING DID NOT IDENTIFY ANY OTHER COMPLAINTS FOR THE GLP CENTRIFUGE MODULE (CM) OR THE PCB ROBOT DISTRIBUTION 5 WITH REGARDS TO THE CUSTOMER REPORTED EVENT. A REVIEW OF THE MANUFACTURING DOCUMENTATION DID NOT IDENTIFY ANY ISSUES AS IT RELATES TO THE CUSTOMER REPORTED EVENT. LABELING WAS REVIEWED AND WAS FOUND TO ADDRESS THE CUSTOMER REPORTED ISSUE. BASED ON THE AVAILABLE INFORMATION, NO SYSTEMIC ISSUE OR DEFICIENCY WAS IDENTIFIED FOR THE GLP CENTRIFUGE MODULE (CM) SERIAL (B)(6) AND PRINTED CIRCUIT BOARD (PCB) ROBOT DISTRIBUTION 5.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THE GLP CENTRIFUGE MODULE HAD A GRIPPER ROBOT ROTATOR ERROR. UPON FURTHER ASSESSMENT, IT WAS NOTED THAT THERE WAS BURNING ON A PRINTED CIRCUIT BOARD (PCB). THERE WAS NO IMPACT TO USER SAFETY AND THERE WAS NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Description of Event or Problem · 0

THE CUSTOMER REPORTED THE GLP CENTRIFUGE MODULE HAD A GRIPPER ROBOT ROTATOR ERROR. UPON FURTHER ASSESSMENT, IT WAS NOTED THAT THERE WAS BURNING ON A PRINTED CIRCUIT BOARD (PCB). THERE WAS NO IMPACT TO USER SAFETY AND THERE WAS NO IMPACT TO PATIENT MANAGEMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
948467 GLP CENTRIFUGE MODULE CALCULATOR/DATA PROCESSING MODULE, FOR CLINICAL USE JQP ABBOTT AUTOMATION SOLUTIONS GMBH 00380740186647

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown