BD LUER-LOK
Report
- Report Number
- 1213809-2025-00461
- Event Type
- Malfunction
- Date Received
- July 10, 2025
- Date of Event
- June 25, 2025
- Report Date
- September 16, 2025
- Manufacturer
- BECTON DICKINSON MEDICAL SYSTEMS
- Product Code
- FMF
- UDI-DI
- 00382903096954
- PMA / PMN Number
- K110771
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CT, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
(B)(4) FOLLOW UP FOR DEVICE EVALUATION: TWENTY-FIVE SAMPLES AND ONE PHOTO OF 10 ML CONTROL SYRINGES (PART NUMBER 309695) FROM BATCH 5121447 WERE RECEIVED AND EVALUATED. TWENTY-ONE SYRINGES WERE FOUND ACCEPTABLE WITH NO DEFECTS OBSERVED. FOUR SYRINGES EXHIBITED BLACK DISCOLORATION ACROSS THE BOTTOM WEB; TWO ALSO HAD LARGE HOLES IN THE WEB, ONE HAD AN OPEN SEAL GREATER THAN ¼ INCH, AND ONE REMAINED SEALED. THE PHOTO CONFIRMED THE PRESENCE OF BLACK DISCOLORATION ON THE AFFECTED PACKAGES. THE CONDITION OBSERVED IS NON-CONFORMING PER PRODUCT SPECIFICATIONS. THE POTENTIAL ROOT CAUSE OF THE PACKAGING DAMAGE AND FOREIGN MATTER DEFECTS IS ASSOCIATED WITH THE PACKAGING PROCESS. INTERVIEWS WITH PRODUCTION PERSONNEL, ALONG WITH A REVIEW OF TECHNICAL AND ELECTRICAL RECORDS AND THE PRODUCTION DATABASE, REVEALED FLUCTUATIONS IN SEAL TEMPERATURE, WHICH MAY HAVE CONTRIBUTED TO THE OBSERVED DEFECTS. FURTHERMORE, A DEVICE HISTORY RECORD REVIEW WAS COMPLETED FOR PROVIDED LOT NUMBER 5121447. A REVIEW SHOWED NO REJECTED INSPECTIONS OR QUALITY ISSUES DURING THE PRODUCTION THAT COULD HAVE CONTRIBUTED TO THE REPORTED DEFECT.
H.3. IF A DEVICE EVALUATION AND/OR DEVICE HISTORY REVIEW IS COMPLETED, A SUPPLEMENTAL REPORT WILL BE FILED.
IT WAS REPORTED THAT THE BD SYRINGE CONTROL 10ML LL PACKAGE WAS DAMAGED / DEFECTIVE. THE FOLLOWING INFORMATION WAS PROVIDED BY THE INITIAL REPORTER: MATERIAL#: 309695, LOT#: 5121447. IT WAS REPORTED BY CUSTOMER THAT ITEM IS BUSTED OPEN INSIDE THE INDIVIDUAL PLASTIC BAGS. RCC RECEIVED A COMPLAINT VIA EMAIL. INJURIES OR ADVERSE EVENT: NO, ITEM: 309695, QUANTITY AFFECTED: 1 BX, SERIAL/LOT NUMBER: (B)(6), PO: (B)(4), ARE ANY SAMPLES AVAILABLE FOR RETURN? YES. REPORTED ISSUE: ITEM IS BUSTED OPEN INSIDE THE INDIVIDUAL PLASTIC BAGS. SHIPPING BOXES ARE NOT DAMAGED AND IT IS ONLY SOME OF THE SYRINGES. CUSTOMER DISPOSITION REQUEST: CREDIT. CUSTOMER RESPONSE ON (B)(6) 2025. THIS ISSUE WAS DISCOVERED ON (B)(6) 2025. AS FOR HOW MANY WERE AFFECTED, MAYBE 5 BUT I WAS UNCOMFORTABLE USING ANYTHING FROM THAT BOX SO I REPLACED ALL OF THEM AND SEALED IT SHUT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 894845 | BD LUER-LOK | PISTON SYRINGE | FMF | BECTON DICKINSON MEDICAL SYSTEMS | 5121447 | 00382903096954 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |