STEALTHSTATION¿ S8 SYSTEM
Report
- Report Number
- 1723170-2025-02671
- Event Type
- Malfunction
- Date Received
- July 10, 2025
- Date of Event
- June 18, 2025
- Report Date
- October 10, 2025
- Manufacturer
- MEDTRONIC NAVIGATION, INC.
- Product Code
- HAW
- UDI-DI
- 00643169722187
- PMA / PMN Number
- K162309
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CO, US
- Reporter Occupation
- 003
Narratives
H2, ADDITIONAL INFORMATION: THE VENDOR ANALYSIS CONFIRMED A BATTERY FAULT WITHIN THE CAMERA. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
CONTINUATION OF D10: SECTION D INFORMATION REFERENCES THE MAIN COMPONENT OF THE SYSTEM. OTHER RELEVANT DEVICE(S) ARE: PRODUCT ID: 9735821R, H3, H6: A MEDTRONIC REPRESENTATIVE WENT TO THE SITE TO TEST THE EQUIPMENT. THE CAMERA WAS REPLACED AND THE SYSTEM THEN PASSED TESTING. CODES B01, C08, AND D02 ARE APPLICABLE TO THIS ANALYSIS. THE CAMERA, LOT NUMBER: P901288, WAS RETURNED TO THE MANUFACTURER FOR ANALYSIS. THE RETURNED POSITIONING SENSOR UNIT (PSU) HAD SCRATCHES ON THE HOUSING AND LENSES. A CHECK OF THE EVENT LOG SHOWED INTERMITTENT FIRMWARE INCOMPATIBILITY DATING BACK TO NOV 2020. THERE WAS A BATTERY VOLTAGE LOW MESSAGE ALONG WITH BUMP DETECTED AND STORAGE TEMPERATURE EXCEEDED. THE PSU FAILED AN ACCURACY TEST (AAK) AT .452MM WITH A PASSING THRESHOLD OF .250MM. THIS PSU HAD NOT BEEN PREVIOUSLY RETURNED FOR A FAILURE. CODES B01, C08, AND D02 ARE APPLICABLE TO THIS ANALYSIS. MEDTRONIC SUBMITS THIS REPORT TO COMPLY WITH FDA REGULATIONS 21 CFR PARTS 4 AND 803. MEDTRONIC HAS MADE REASONABLE EFFORTS TO PROVIDE AS MUCH RELEVANT INFORMATION AS IS AVAILABLE TO THE COMPANY AS OF THE SUBMISSION DATE OF THIS REPORT. THIS REPORT DOES NOT CONSTITUTE AN ADMISSION OR A CONCLUSION BY FDA, MEDTRONIC, OR ITS EMPLOYEES THAT THE DEVICE, MEDTRONIC, OR ITS EMPLOYEE CAUSED OR CONTRIBUTED TO THE EVENT DESCRIBED IN THE REPORT. ANY REQUIRED FIELDS THAT ARE UNPOPULATED ARE BLANK BECAUSE THE INFORMATION IS CURRENTLY UNKNOWN OR UNAVAILABLE. MEDTRONIC WILL SUBMIT A SUPPLEMENTAL REPORT IF ADDITIONAL RELEVANT INFORMATION BECOMES KNOWN.
MEDTRONIC RECEIVED INFORMATION REGARDING A NAVIGATION SYSTEM BEING USED OUTSIDE OF A PROCEDURE. IT WAS REPORTED THAT A LOCALIZER FAULTED ERROR APPEARED ON THE CAMERA WITH GREEN AND AMBER LIGHTS. IT WAS NOTED THAT THE NETWORK DEVICE INTERFACE (NDI) TOOL-BOX SCREEN SHOWED STORAGE TEMPERATURE AND BUMP STATUS HIGHLIGHTED. THIS ISSUE WAS NOTICED DURING SET UP FOR A CASE AND THE SYSTEM WAS THEN SWAPPED OUT. THERE WAS NO PATIENT PRESENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1602102 | STEALTHSTATION¿ S8 SYSTEM | NEUROLOGICAL STEREOTAXIC INSTRUMENT | HAW | MEDTRONIC NAVIGATION, INC. | 9735665 | 00643169722187 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |