FDA Adverse Event Malfunction Summary report: N

TMINI MINIATURE ROBOTIC SYSTEM

MDR report key: 22469879 · Received July 10, 2025

Report

Report Number
3000719653-2025-00007
Event Type
Malfunction
Date Received
July 10, 2025
Date of Event
June 10, 2025
Report Date
July 10, 2025
Manufacturer
THINK SURGICAL, INC
Product Code
OLO
UDI-DI
M9461090910
PMA / PMN Number
K250877
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 0

UPON COMPLETION OF A ROBOTIC PROCEDURE, THE HOSPITAL REPORTED THAT THE PLASTIC FRICTION WASHER OF THE ARRAY BRACKET WAS FOUND TO BE DAMAGED AND HAD DETACHED DURING DISASSEMBLY AND TRANSFER OUT OF THE STERILE FIELD. THE PROCEDURE WAS COMPLETED WITHOUT INCIDENT, AND NO HARM WAS REPORTED TO THE PATIENT AND STAFF. THE INTERNAL INVESTIGATION OF THE ARRAY BRACKET SHOWED THAT THE WASHER WAS MISSING AS IT WAS REPORTED AND CONFIRMED BY THINK SURGICAL CLINICAL REPRESENTATIVE. THIS EVENT IS BEING REPORTED AS A MALFUNCTION, AS THE WASHER FAILED STRUCTURALLY. ALTHOUGH NO HARM OCCURRED IN THIS INSTANCE, THERE IS A POTENTIAL THAT THE WASHER COULD DETACH AND FALL INTO THE STERILE FIELD OR SURGICAL SITE DURING A PROCEDURE.

Description of Event or Problem · 0

UPON CONCLUSION OF A SUCCESSFULLY COMPLETED ROBOTIC CASE, THE STAFF NOTICED THE PLASTIC FRICTION RING OF THE ARRAY BRACKET WAS DAMAGED AND COMING OFF THE BRACKET DURING DISASSEMBLY AND TRANSFER TO THE BACK TABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
978988 TMINI MINIATURE ROBOTIC SYSTEM ORTHOPEDIC STEREOTAXIC INSTRUMENT OLO THINK SURGICAL, INC 109091 22068 M9461090910

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown