TMINI MINIATURE ROBOTIC SYSTEM
Report
- Report Number
- 3000719653-2025-00008
- Event Type
- Malfunction
- Date Received
- July 10, 2025
- Date of Event
- June 17, 2025
- Report Date
- July 10, 2025
- Manufacturer
- THINK SURGICAL, INC
- Product Code
- OLO
- UDI-DI
- M9461090910
- PMA / PMN Number
- K250877
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO, US
- Reporter Occupation
- OTHER
Narratives
UPON COMPLETION OF A ROBOTIC SURGICAL PROCEDURE, STAFF OBSERVED THAT THE PLASTIC FRICTION WASHER ON THE ARRAY BRACKET WAS LOOSE. THE PROCEDURE WAS COMPLETED SUCCESSFULLY, AND NO HARM TO THE PATIENT OR STAFF WAS REPORTED. TI INVESTIGATION DETERMINED THAT THE FRICTION WASHER WAS BROKEN, BUT STILL IN PLACE ON THE BRACKET. THE DEVICE FUNCTIONED AS INTENDED DURING THE SURGERY. THIS EVENT IS BEING REPORTED AS A MALFUNCTION, AS THE WASHER FAILED STRUCTURALLY. ALTHOUGH NO HARM OCCURRED IN THIS INSTANCE, THERE IS A POTENTIAL THAT THE WASHER COULD DETACH AND FALL INTO THE STERILE FIELD OR SURGICAL SITE DURING A PROCEDURE.
UPON CONCLUSION OF A SUCCESSFULLY COMPLETED ROBOTIC CASE, THE STAFF NOTICED THE PLASTIC FRICTION RING OF THE ARRAY BRACKET WAS LOOSE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 978985 | TMINI MINIATURE ROBOTIC SYSTEM | ORTHOPEDIC STEREOTAXIC INSTRUMENT | OLO | THINK SURGICAL, INC | 109091 | 22068 | M9461090910 |