FDA Adverse Event Malfunction Summary report: N

BALLARD CLOSED SUCTION SYSTEM

MDR report key: 22469275 · Received July 10, 2025

Report

Report Number
8030647-2025-00066
Event Type
Malfunction
Date Received
July 10, 2025
Date of Event
June 9, 2025
Report Date
November 4, 2025
Manufacturer
AVANOS MEDICAL INC.
Product Code
BSY
UDI-DI
00609038983103
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

ADDITIONAL INFORMATION-INVESTIGATION COMPLETION: D4; H2; H6. THE PRODUCT IN QUESTION WAS NOT RETURNED FOR EVALUATION; HOWEVER, THE REPORTER DID PROVIDE PHOTOGRAPHIC EVIDENCE OF THE DEVICE IN QUESTION; ANALYSIS OF THE PHOTOGRAPH CONFIRMED THE COMPLAINT AS REPORTED. WITHOUT A SAMPLE TO PERFORM FUNCTIONAL EVALUATION, A ROOT CAUSE COULD NOT BE DETERMINED. THE DEVICE HISTORY RECORD FOR LOT 1581059 WAS REVIEWED AND THE PRODUCT WAS PRODUCED ACCORDING TO PRODUCT SPECIFICATIONS. ALL INFORMATION REASONABLY KNOWN AS OF 04 NOV 2025 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY SUN MED HOLDINGS LLC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO SUN MED HOLDINGS LLC. SUN MED HOLDINGS LLC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE SUN MED HOLDINGS COMPLAINT DATABASE AND IDENTIFIED AS (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN SUN MED HOLDINGS LLC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.

Additional Manufacturer Narrative · 0

THE PRODUCT INVOLVED IN THE REPORT HAS NOT BEEN RETURNED FOR EVALUATION. A REVIEW OF THE DEVICE HISTORY RECORD IS IN-PROGRESS. ALL INFORMATION REASONABLY KNOWN AS OF 10 JUL 2025 HAS BEEN INCLUDED IN THIS HEALTH AUTHORITY REPORT. SHOULD ADDITIONAL INFORMATION BE OBTAINED, A FOLLOW-UP HEALTH AUTHORITY REPORT WILL BE PROVIDED. THE INFORMATION PROVIDED BY SUN MED HOLDINGS LLC. REPRESENTS ALL OF THE KNOWN INFORMATION AT THIS TIME. DESPITE GOOD FAITH EFFORTS TO OBTAIN ADDITIONAL INFORMATION, THE COMPLAINANT / REPORTER WAS UNABLE OR UNWILLING TO PROVIDE ANY FURTHER PATIENT, PRODUCT, OR PROCEDURAL DETAILS TO SUN MED HOLDINGS LLC. SUN MED HOLDINGS LLC. HAS NO INDEPENDENT KNOWLEDGE OF THE EVENT REPORTED BUT IS RELAYING THE INFORMATION THAT WAS PROVIDED BY THE USER FACILITY WHERE THE INCIDENT OCCURRED. THIS PRODUCT INCIDENT IS DOCUMENTED IN THE SUN MED HOLDINGS COMPLAINT DATABASE AND IDENTIFIED AS (B)(4). THIS INFORMATION IS SUBMITTED PURSUANT TO 21CFR803, IN COMPLIANCE WITH THE MEDICAL DEVICE REPORTING REQUIREMENT AND SHOULD NOT BE CONSIDERED TO BE AN ADMISSION THAT AN SUN MED HOLDINGS LLC. PRODUCT IS DEFECTIVE OR CAUSED SERIOUS INJURY.

Description of Event or Problem · 0

SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE FIRST OF TWO REPORTS. REFER TO 8030647-2025-00067 FOR THE SECOND REPORT. IT WAS REPORTED, WHEN THE REGISTERED NURSE ATTEMPTED TO FLUSH THE CATHETER AFTER SUCTIONING THE FLUSH FLUID WENT INTO THE CLEAR PROTECTIVE SLEEVE INSTEAD OF THE CATHETER; THERE WAS NO REPORTED INJURY. THE CATHETER WAS REPLACED.

Description of Event or Problem · 0

SUNMED HOLDINGS LLC. RECEIVED A SINGLE REPORT THAT REFERENCED TWO DIFFERENT INCIDENCES, WHICH WERE ASSOCIATED WITH SEPARATE UNITS, INVOLVING THE SAME PATIENT. THIS IS THE FIRST OF TWO REPORTS. REFER TO 8030647-2025-00067 FOR THE SECOND REPORT. IT WAS REPORTED, WHEN THE REGISTERED NURSE ATTEMPTED TO FLUSH THE CATHETER AFTER SUCTIONING THE FLUSH FLUID WENT INTO THE CLEAR PROTECTIVE SLEEVE INSTEAD OF THE CATHETER; THERE WAS NO REPORTED INJURY. THE CATHETER WAS REPLACED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1602055 BALLARD CLOSED SUCTION SYSTEM BALLARD CLOSED SUCTION CATHETER, DOUBLE SWIVEL ELBOW BSY AVANOS MEDICAL INC. 98310 1581059 00609038983103

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown