STYLE 115 SILICONE GEL FILLED BREAST IMPLANT
Report
- Report Number
- 9617229-2025-11496
- Event Type
- Injury
- Date Received
- July 10, 2025
- Date of Event
- May 22, 2025
- Report Date
- July 10, 2025
- Manufacturer
- ALLERGAN (COSTA RICA)
- Product Code
- FTR
- PMA / PMN Number
- P020056
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- KS
- Reporter Occupation
- PHYSICIAN
- Health Professional
- Yes
Narratives
CONTINUED: E.1. PHONE NUMBER: (B)(6). A REVIEW OF THE DEVICE HISTORY RECORD HAS BEEN COMPLETED. NO DEVIATIONS OR NON-CONFORMANCES NOTED. FURTHER INFORMATION FROM THE REPORTER REGARDING EVENT, PRODUCT, OR PATIENT DETAILS HAS BEEN REQUESTED. NO ADDITIONAL INFORMATION IS AVAILABLE AT THIS TIME. REASON FOR REOPERATION: RUPTURE.
ADDITIONAL, CHANGED, AND/OR CORRECTED DATA: D9, H3, H6. DEVICE EVALUATION: THE DEVICE RELATED TO THE REPORTED EVENTS RUPTURE WAS RECEIVED ON (B)(6) 2025 WITH LOT NUMBER 2150713. BASED ON THE PRODUCT ANALYSIS PERFORMED, THE ASSESSMENTS OF THE COMPLAINT CODES ARE: RUPTURE: OBSERVED AN OPENING ASSESSED AS SURGICAL IMPACT OPENING. AS PER THE INVESTIGATION PROCEDURE, CREASES AND NON PENETRATING NICKS WERE OBSERVED AND NONE OF THE OBSERVATIONS ARE FOUND TO BE POTENTIALLY RELATED TO THE MANUFACTURING PROCESS, NO FURTHER ACTIONS ARE REQUIRED.
HEALTHCARE PROFESSIONAL REPORTED LEFT SIDE RUPTURE. DEVICE HAS BEEN EXPLANTED.
HEALTHCARE PROFESSIONAL REPORTED RUPTURE CONFIRMED BY ULTRASOUND. THIS RECORD IS FOR THE LEFT SIDE. DEVICE WAS EXPLANTED AND REPLACED WITH A NON-ABBVIE DEVICE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1729813 | STYLE 115 SILICONE GEL FILLED BREAST IMPLANT | PROSTHESIS, BREAST, NONINFLATABLE, INTERNAL, SILICONE GEL-FILLED | FTR | ALLERGAN (COSTA RICA) | 2150713 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 43 YR | Female | Required Intervention |