TRUSTEEL
Report
- Report Number
- 3003442380-2025-11467
- Event Type
- Malfunction
- Date Received
- July 10, 2025
- Date of Event
- June 14, 2025
- Report Date
- September 17, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244018426
- PMA / PMN Number
- K041545
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
INITIAL AND FINAL MDR (B)(4) - DEVICE 2 OF 2.
SUPPLEMENTAL REPORT 01 - (B)(4) - MDR 3003442380-2025-11467. ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: REVISION 21 OF (B)(4) DOES NOT REQUIRE A COMPLAINT THAT IS TYPE 2 REPORTABLE TO OPEN A CHILD INVESTIGATION. THIS CHILD INVESTIGATION WAS OPENED AGAINST A PREVIOUS REVISION OF (B)(4). COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6012370, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6012370 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 101 AND PACKAGING IN THE MACHINE MULTIVAC 14 ON 17-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING TUBE OF THE LOT 5C00910 WAS MANUFACTURED ACCORDING TO THE WI VERSION 67 AND MANUFACTURED IN THE MACHINE SC05-SC06, ON 16-MAR-2025, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED TWO INFUSION SETS SPLIT/ RIPPED TUBING EVENT ON (B)(6) 2025. INFUSION SETS WERE USED FOR ABOUT 12 HOURS. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 773801 | TRUSTEEL | UNO CONTACT DETACH G29 60/6TCAP 10PK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 1002833 | 6012370 | 05705244018426 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 33 YR | Female |