FDA Adverse Event Injury Summary report: N

DXTEND HIGHMO PE CUP D38 +3MM

MDR report key: 22462428 · Received July 9, 2025

Report

Report Number
1818910-2025-11735
Event Type
Injury
Date Received
July 9, 2025
Date of Event
June 24, 2025
Manufacturer
DEPUY IRELAND - 9616671
Product Code
PHX
UDI-DI
10603295272007
PMA / PMN Number
K073676
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
AS
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL
Health Professional
Yes

Narratives

Additional Manufacturer Narrative · 0

THIS REPORT IS BEING SUBMITTED PURSUANT TO THE PROVISIONS OF 21 CFR, PART 803 (AND/OR PART 4, AS APPLICABLE). THIS REPORT MAY BE BASED ON INFORMATION WHICH HAS NOT BEEN INVESTIGATED OR VERIFIED PRIOR TO THE REQUIRED REPORTING DATE. THIS REPORT DOES NOT REFLECT A CONCLUSION BY DEPUY SYNTHES, OR ITS EMPLOYEES THAT THE REPORT CONSTITUTES AN ADMISSION THAT THE PRODUCT, DEPUY SYNTHES, OR ITS EMPLOYEES CAUSED OR CONTRIBUTED TO THE POTENTIAL EVENT DESCRIBED IN THIS REPORT. IF INFORMATION IS OBTAINED THAT WAS NOT AVAILABLE FOR THE INITIAL REPORT, A FOLLOW-UP REPORT WILL BE FILED AS APPROPRIATE.

Additional Manufacturer Narrative · 0

PRODUCT COMPLAINT # (B)(4). INVESTIGATION SUMMARY: REMOVAL OF INFECTED DELTA XTEND. THE PRODUCT WAS NOT RETURNED TO J&J MEDTECH ORTHOPAEDICS, HOWEVER PHOTOS WERE PROVIDED FOR REVIEW. THE PHOTO INVESTIGATION REVEALED THAT DXTEND HIGHMO PE CUP D38 +3MM EXHIBITS SIGNS OF WEAR ALONG OF THE ARTICULATION SURFACE. THE MODE OF FAILURE OF THE DEVICE IS MULTI-FACTORIAL AND CONSIDERATION MUST BE GIVEN TO ALL OTHER POTENTIAL INFLUENCES SUCH AS SURGICAL PROCESS, PATIENT VARIABLES (E.G. ACTIVITY LEVEL AND USE), ANATOMICAL CONSIDERATIONS, PATIENT CHANGES OVER TIME AND EXTRACTION PROCESS. ANY CONCLUSIONS FROM THE INVESTIGATIONAL INPUT PROVIDED HAVE TO BE PLACED INTO CONTEXT WITH ALL OTHER RELEVANT FACTORS. A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE 130738003 LOT NUMBER 5232424, AND NO NON-CONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED DURING MANUFACTURING. AS THE DEVICE WAS NOT RETURNED, AN AS-RECEIVED CONDITION COULD NOT BE ASSESSED, AND A DIMENSIONAL INSPECTION AND DOCUMENT/SPECIFICATION REVIEW WERE NOT COMPLETED. THE OVERALL COMPLAINT WAS NOT CONFIRMED AS THE OBSERVED CONDITION OF THE DXTEND HIGHMO PE CUP D38 +3MM WOULD HAVE NOT CONTRIBUTED TO THE REPORTED ADVERSE EVENT. BASED ON THE INVESTIGATION FINDINGS, IT HAS BEEN DETERMINED THAT NO CORRECTIVE AND/OR PREVENTATIVE ACTION IS PROPOSED. THERE IS NO INDICATION THAT A DESIGN OR MANUFACTURING ISSUE HAS CAUSED THE REPORTED COMPLAINT CONDITION. AS PART OF J&J MEDTECH ORTHOPAEDICS QUALITY PROCESS, ALL DEVICES ARE MANUFACTURED, INSPECTED, AND RELEASED TO APPROVED SPECIFICATIONS. ADDITIONAL MONITORING FOR ANY POTENTIAL SAFETY SIGNALS WILL BE CONDUCTED THROUGH COMPLAINT TRENDING AND OTHER POST-MARKET SAFETY SURVEILLANCE ACTIVITIES. DEVICE HISTORY LOT: A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE 130738003 LOT NUMBER 5232424, AND NO NON-CONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED DURING MANUFACTURING. DEVICE HISTORY REVIEW: A MANUFACTURING RECORD EVALUATION WAS PERFORMED FOR THE FINISHED DEVICE PRODUCT CODE 130738003 LOT NUMBER 5232424, AND NO NON-CONFORMANCES / MANUFACTURING IRREGULARITIES WERE IDENTIFIED DURING MANUFACTURING. H11 ADDITIONAL NARRATIVE: ADDED: D10.

Description of Event or Problem · 0

IT WAS REPORTED THAT PATIENT UNDERWENT REMOVAL OF DELTA XTEND DUE TO INFECTION. DOI: UNKNOWN 2015, DOR: (B)(6) 2025, AFFECTED SIDE: UNKNOWN.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
601155 DXTEND HIGHMO PE CUP D38 +3MM SHOULDER HUMERAL CUP PHX DEPUY IRELAND - 9616671 5232424 10603295272007

Patients

Seq Age Sex Outcome Treatment
1 78 YR Female Required Intervention