AUTOSOFT XC
Report
- Report Number
- 3003442380-2025-11407
- Event Type
- Malfunction
- Date Received
- July 9, 2025
- Date of Event
- June 10, 2025
- Report Date
- September 2, 2025
- Manufacturer
- UNOMEDICAL DEVICES S.A. DE C.V.
- Product Code
- FPA
- UDI-DI
- 05705244016620
- PMA / PMN Number
- K032854
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- CA, US
- Reporter Occupation
- 003
Narratives
ADDITIONAL INFORMATION - THIS MDR IS BEING SUBMITTED TO INCLUDE THE BELOW: H6: INVESTIGATION RESULTS UNDER TYPE OF INVESTIGATION, INVESTIGATION FINDINGS, INVESTIGATION CONCLUSIONS. H11: INVESTIGATION SUMMARY: REVISION 21 OF (B)(4) DOES NOT REQUIRE A COMPLAINT THAT IS TYPE 2 REPORTABLE TO OPEN A CHILD INVESTIGATION. THIS CHILD INVESTIGATION WAS OPENED AGAINST A PREVIOUS REVISION OF (B)(4). COMPLAINT INVESTIGATION RESULTS: A COMPLAINT INVESTIGATION HAS BEEN INITIATED UNDER COMPLAINT INVESTIGATION CHILD RECORD (B)(4). THE BATCH 6010720, IN QUESTION WAS MANUFACTURED AT THE REYNOSA SITE. DEVICE HISTORY RECORD (DHR) REVIEW: THE LOT 6010720 WAS MANUFACTURED ACCORDING TO THE WORK INSTRUCTION (WI) VERSION 120 AND MANUFACTURED IN THE LINE 4, ON 14/DEC/2024, WITH A TOTAL OF (B)(4) UNITS. THE SUB-ASSEMBLY, GLUING OF TUBING OF THE LOT 4M01053 WAS MANUFACTURED ACCORDING TO THE WI VERSION 12 AND MANUFACTURED IN THE MACHINE IGTL01, ON 09/DEC/2024, WITH A TOTAL OF (B)(4) UNITS. REVIEW OF THE DHR SHOWED THAT ALL RELEVANT TESTS REQUIRED DURING THE RELATED PROCESSES HAD BEEN FULFILLED AND MET THE REQUIREMENTS. NO DEVIATION WAS IDENTIFIED, NOR MAINTENANCE EVENTS WERE RECORDED RELATED TO COMPLAINT CODE. CONCLUSION SUMMARY OF COMPLAINT INVESTIGATION: AS A RESULT OF THE FOLLOWING: NO NON-CONFORMANCE (NC) RAISED DURING PRODUCTION RELATED TO COMPLAINT CODE, NO FURTHER ACTIONS ARE REQUIRED. THIS COMPLAINT WILL NOT REQUIRE FURTHER ROOT CAUSE INVESTIGATION NOR CORRECTIVE AND PREVENTIVE ACTION (CAPA) PLAN. THEREFORE, THIS ISSUE WILL BE MONITORED THROUGH THE POST MARKET SURVEILLANCE ACTIVITIES.
REFERENCE NUMBER (B)(4). EVENT OCCURRED IN THE UNITED STATES. IT WAS REPORTED THAT THE PATIENT FACED AN INFUSION SET TUBING DETACHED EVENT ON (B)(6) 2025 FROM CONNECTOR. INFUSION SET WAS USED FOR TWO DAYS. PATIENT REPLACED INFUSION SET AND RESUMED INSULIN DELIVERIES SUCCESSFULLY. NO FURTHER INFORMATION AVAILABLE.
TO DATE NO ADDITIONAL PATIENT OR EVENT DETAILS HAVE BEEN RECEIVED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 577172 | AUTOSOFT XC | UNO INSET I 60/6 GREY TCAP 10PK INT | FPA | UNOMEDICAL DEVICES S.A. DE C.V. | 1001680 | 6010720 | 05705244016620 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 41 YR | Male |