FDA Adverse Event Malfunction Summary report: N

FISHER & PAYKEL HEALTHCARE

MDR report key: 22454709 · Received July 8, 2025

Report

Report Number
9611451-2025-00634
Event Type
Malfunction
Date Received
July 8, 2025
Report Date
November 27, 2025
Manufacturer
FISHER & PAYKEL HEALTHCARE LTD
Product Code
QAV
UDI-DI
09420012466662
PMA / PMN Number
K221338
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
OTHER
Health Professional
N

Narratives

Additional Manufacturer Narrative · 0

(B)(6). FISHER & PAYKEL HEALTHCARE (F&P) HAS REQUESTED FOR THE COMPLAINT DEVICE TO BE RETURNED FOR EVALUATION. F&P WILL PROVIDE A FOLLOW UP REPORT UPON COMPLETION OF F&P'S INVESTIGATION. PRODUCT BACKGROUND: THE AIRVO 3 IS FOR THE TREATMENT OF SPONTANEOUSLY BREATHING PATIENTS WHO WOULD BENEFIT FROM RECEIVING HIGH FLOW WARMED AND HUMIDIFIED RESPIRATORY GASES. THIS INCLUDES PATIENTS WHO HAVE HAD UPPER AIRWAYS BYPASSED. THE FLOW MAY BE FROM 2 70 L/MIN DEPENDING ON THE PATIENT INTERFACE. THE AIRVO 3 IS FOR PATIENTS IN HOSPITALS AND SUBACUTE FACILITIES. THE AIRVO 3 IS NOT INTENDED FOR LIFE SUPPORT. THE AIRVO 3 CAN DELIVER THESE HIGH FLOW GASES THROUGH NASAL CANNULA TO AUGMENT THE BREATHING OF SPONTANEOUSLY BREATHING NEONATE, INFANT, CHILD, ADOLESCENT AND ADULT PATIENTS SUFFERING FROM RESPIRATORY DISTRESS AND/OR HYPOXEMIA IN THE HOSPITAL SETTING. THE AIRVO 3 IS NOT INTENDED TO PROVIDE TOTAL VENTILATORY REQUIREMENTS OF THE PATIENT AND IS NOT FOR USE DURING FIELD TRANSPORT.

Additional Manufacturer Narrative · 0

(B)(4). CORRECTED DATA: B4, B5, D9, G3, G6, H2, H3, H6, H11. PRODUCT BACKGROUND: THE AIRVO 3 IS FOR THE TREATMENT OF SPONTANEOUSLY BREATHING PATIENTS WHO WOULD BENEFIT FROM RECEIVING HIGH FLOW WARMED AND HUMIDIFIED RESPIRATORY GASES. THIS INCLUDES PATIENTS WHO HAVE HAD UPPER AIRWAYS BYPASSED. THE FLOW MAY BE FROM 2 - 70 L/MIN DEPENDING ON THE PATIENT INTERFACE. THE AIRVO 3 IS FOR PATIENTS IN HOSPITALS AND SUBACUTE FACILITIES. THE AIRVO 3 IS NOT INTENDED FOR LIFE SUPPORT. THE AIRVO 3 CAN DELIVER THESE HIGH FLOW GASES THROUGH NASAL CANNULA TO AUGMENT THE BREATHING OF SPONTANEOUSLY BREATHING NEONATE, INFANT, CHILD, ADOLESCENT AND ADULT PATIENTS SUFFERING FROM RESPIRATORY DISTRESS AND/OR HYPOXEMIA IN THE HOSPITAL SETTING. THE AIRVO 3 IS NOT INTENDED TO PROVIDE TOTAL VENTILATORY REQUIREMENTS OF THE PATIENT AND IS NOT FOR USE DURING FIELD TRANSPORT. METHOD: THE SUBJECT AIRVO 3 WAS RECEIVED BY FISHER & PAYKEL HEALTHCARE (F&P) WHERE IT WAS INSPECTED BY A TRAINED F&P INVESTIGATION ENGINEER. THE DEVICE WAS PERFORMANCE TESTED AND THE AUDIBLE ALARM FUNCTION WAS CHECKED. F&P'S INVESTIGATION IS BASED ON THE INFORMATION OBTAINED FROM THE HEALTHCARE FACILITY, F&P'S EVALUATION OF THE SUBJECT DEVICE, PREVIOUS INVESTIGATIONS OF SIMILAR COMPLAINTS, AND F&P'S KNOWLEDGE OF THE PRODUCT. RESULTS: DURING TESTING IT WAS FOUND THAT THE AIRVO 3 DID NOT HAVE AN AUDIBLE ALARM, CONFIRMING THE REPORTED EVENT. IT WAS IDENTIFIED THAT THE ISSUE WAS DUE TO A FAULTY SPEAKER. THE SPEAKER IS A SUPPLIED COMPONENT THAT IS ASSEMBLED INTO THE AIRVO 3. THERE WAS NO PATIENT INVOLVEMENT REPORTED BY THE HEALTHCARE FACILITY. CONCLUSION: F&P'S INVESTIGATION WAS UNABLE TO DETERMINE THE EXACT CAUSE OF THE REPORTED EVENT. BASED ON THE INFORMATION PROVIDED BY THE HEALTHCARE FACILITY, F&P'S INVESTIGATION, AND KNOWLEDGE OF THE PRODUCT, A POSSIBLE CAUSE OF THE SPEAKER FAULT WAS DUE TO THE SPEAKER COMPONENT BEING OUT OF SPECIFICATION. THE AIRVO 3 WILL GENERATE A VISUAL ALARM IN THE PRESENCE OF AN INAUDIBLE ALARM. THE ALARM IN THE REPORTED EVENT PREVENTS THE OPERATION OF THE SUBJECT DEVICE AND ALERTS THE USER TO TAKE APPROPRIATE AND TIMELY ACTION. THERE WAS NO PATIENT INVOLVEMENT REPORTED BY THE HEALTHCARE FACILITY. THE USER INSTRUCTIONS WHICH ACCOMPANY THE AIRVO 3 PROVIDE GUIDANCE ON TROUBLESHOOTING ALARMS GENERATED BY THE DEVICE. THE USER INSTRUCTIONS STATE THE FOLLOWING: - DEVICE FAULT (FAULT X.X.X): "A TECHNICAL FAULT HAS OCCURRED, AND THE PATIENT MAY NEED PROMPT ATTENTION. RESTART THE DEVICE TO CLEAR THE FAULT CONDITION. IF THE ISSUE PERSISTS, CONTACT YOUR SERVICE REPRESENTATIVE." THE USER INSTRUCTIONS ALSO STATE: "MAKE SURE THE AUDITORY ALARM SIGNAL IS AUDIBLE TO THE OPERATOR WHO WILL RESPOND TO ALARMS BY FOLLOWING THE INSTRUCTIONS IN SECTION 7.5 TO TEST THE ALARM BEFORE STARTING THERAPY." "DO NOT USE THE AIRVO 3 IF IT FAILS [THE 7.5 CHECKING THE ALARM SYSTEM] TEST. CONTACT YOUR FISHER & PAYKEL HEALTHCARE REPRESENTATIVE." "THE AIRVO 3 IS FOR THE TREATMENT OF SPONTANEOUSLY BREATHING PATIENTS WHO WOULD BENEFIT FROM RECEIVING HIGH FLOW WARMED AND HUMIDIFIED RESPIRATORY GASES. THIS INCLUDES PATIENTS WHO HAVE HAD UPPER AIRWAY BYPASSED." "THE AIRVO 3 IS NOT INTENDED TO PROVIDE TOTAL VENTILATORY REQUIREMENTS OF THE PATIENT AND IS NOT FOR USE DURING FIELD TRANSPORT." "THE AIRVO 3 IS NOT INTENDED FOR LIFE SUPPORT. DO NOT USE AIRVO 3 ON PATIENTS WHO CANNOT TOLERATE A BRIEF INTERRUPTION OF THERAPY." "APPROPRIATE PATIENT MONITORING IS REQUIRED FOR ALL PATIENTS USING THE AIRVO 3."

Description of Event or Problem · 0

A HEALTHCARE FACILITY IN NEW JERSEY REPORTED VIA A FISHER AND PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT A PT301 AIRVO 3 DID NOT HAVE AN AUDIBLE ALARM. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

Description of Event or Problem · 0

A HEALTHCARE FACILITY IN NEW JERSEY REPORTED VIA A FISHER AND PAYKEL HEALTHCARE (F&P) FIELD REPRESENTATIVE THAT A PT301 AIRVO 3 DID NOT HAVE AN AUDIBLE ALARM, AND THAT THE PT301 AIRVO 3 GENERATED A VISUAL ALARM. THE NATURE OF THE VISUAL ALARM SUGGESTS THERE MAY BE A POTENTIAL ISSUE WITH THE SUBJECT DEVICE'S AUDIBLE ALARM. THERE WAS NO REPORTED PATIENT INVOLVEMENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
451170 FISHER & PAYKEL HEALTHCARE RESPIRATORY HUMIDIFIER QAV FISHER & PAYKEL HEALTHCARE LTD PT301US 2102841492 09420012466662

Patients

Seq Age Sex Outcome Treatment
1 NA Unknown