FDA Adverse Event
Malfunction
Summary report: N
WARMTOUCH COMPACT WARMING UNIT
MDR report key: 2245014
·
Received July 29, 2011
Report
- Report Number
- 2936999-2011-00509
- Event Type
- Malfunction
- Date Received
- July 29, 2011
- Date of Event
- June 29, 2011
- Report Date
- June 29, 2011
- Manufacturer
- COVIDIEN/FORMERLY TYCO
- Product Code
- DWJ
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- UK
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
COVIDIEN TECHNICIAN CONFIRMED CASE HAS MELTED IN AREA ABOVE HEATING ELEMENT (AIR INLET AREA). ON TESTING FOUND THAT BLOWER HAS FAILED, LEADING TO ELEMENT OVERHEATING. APPROXIMATE YEAR OF MANUFACTURE IS 1999. THE WT-5800 WARMTOUCH IS NOT DISTRIBUTED IN THE U.S.; HOWEVER, THE WT-5300 WARMTOUCH IS OF ESSENTIALLY IDENTICAL DESIGN AND IS DISTRIBUTED IN THE U.S. THE 510K NUMBER FOR THE WT-5300 WARMTOUCH IS K020604.
Description of Event or Problem · 1
WARMTOUCH WAS RETURNED FOR CHECK FOR PROPER OPERATION. EVALUATION CONFIRMED CASE HAS MELTED IN AREA ABOVE HEATING ELEMENT (AIR INLET AREA). ON TESTING FOUND THAT BLOWER HAS FAILED, LEADING TO ELEMENT OVERHEATING. NO PATIENT WAS INJURED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | WARMTOUCH COMPACT WARMING UNIT | CONVECTIVE AIR WARMER | DWJ | COVIDIEN/FORMERLY TYCO |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |