FDA Adverse Event Malfunction Summary report: N

FULL FACE MASK

MDR report key: 2244662 · Received September 13, 2011

Report

Report Number
9611451-2011-00559
Event Type
Malfunction
Date Received
September 13, 2011
Date of Event
August 16, 2011
Report Date
August 18, 2011
Manufacturer
FISHER & PAYKEL HEALTHCARE LIMITED
Product Code
MNT
PMA / PMN Number
K060044
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TW
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). METHOD: THE RETURNED RT040M FULL FACE MASKS WERE VISUALLY INSPECTED FOR SEAL SEPARATION. RESULTS: VISUAL INSPECTION OF THE RETURNED MASKS REVEALED THAT THE SEALS WERE COMPLETELY SEPARATED FROM THE BASES, CONFIRMING THE REPORTED FAULT. FURTHER INSPECTION SHOWED THAT THE SEALS WERE GLUED PROPERLY WITH A CONTINUOUS BEAD OF GLUE. INDENTS IN THE GLUE LINE SHOW THAT THE SEAL WAS PROPERLY INSERTED INTO THE BASE. A LOT CHECK REVEALED ONE OTHER COMPLAINT OF THIS NATURE FOR LOT NUMBER 090807. CONCLUSION: THE RT040 FULL FACE MASK FEATURES A MASK BASE, SEAL AND HEADGEAR. THE MASK SEAL IS INSERTED INTO THE MASK BASE AND IS SECURED WITH GLUE. THE RT040 MASK IS SUBJECTED TO POST-PRODUCTION TESTS TO ENSURE THE SEAL ON THE MASK CAN WITHSTAND A REMOVAL FORCE GREATER THAN 100N. WE ARE UNABLE TO DETERMINE THE ROOT CAUSE OF THE SEPARATION. VISUAL INSPECTION OF THE RETURNED DEVICE SHOWED THAT THE MASK WAS ASSEMBLED AND GLUED PROPERLY. THIS SUGGESTS THAT THE SEPARATION OCCURRED POST-PRODUCTION, POSSIBLY DUE TO THE DRYING OUT OR SEPARATION OF THE GLUE CAUSED BY ADVERSE CONDITIONS DURING TRANSPORT OR STORAGE. (B)(4).

Description of Event or Problem · 1

A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT THE SILICONE SEALS OF TWO RT040M FULL FACE MASKS WERE FOUND SEPARATED FROM THE MASK BASES, CAUSING IT TO LEAK. THIS WAS NOTICED DURING USE ON A PATIENT. NO PATIENT CONSEQUENCE WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 FULL FACE MASK MNT MNT FISHER & PAYKEL HEALTHCARE LIMITED RT040M 090807

Patients

Seq Age Sex Outcome Treatment
1