FULL FACE MASK
Report
- Report Number
- 9611451-2011-00557
- Event Type
- Malfunction
- Date Received
- September 13, 2011
- Date of Event
- August 16, 2011
- Report Date
- August 18, 2011
- Manufacturer
- FISHER & PAYKEL HEALTHCARE LIMITED
- Product Code
- MNT
- PMA / PMN Number
- K060044
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TW
- Reporter Occupation
- OTHER
Narratives
(B)(4). THE RETURNED RT040 FULL FACE MASK WAS VISUALLY INSPECTED FOR SEAL SEPARATION. VISUAL INSPECTION OF THE RETURNED MASK REVEALED THAT THE SEAL WAS COMPLETELY SEPARATED FROM THE BASE, CONFIRMING THE REPORTED FAULT. FURTHER INSPECTION SHOWED THAT THE SEAL WAS GLUED PROPERLY WITH A CONTINUOUS BEAD OF GLUE. INDENTS IN THE GLUE LINE SHOW THAT THE SEAL WAS PROPERLY INSERTED INTO THE BASE. A LOT CHECK REVEALED NO OTHER COMPLAINTS OF THIS NATURE FOR LOT NUMBER 090807. THE RT040 FULL FACE MASK FEATURES A MASK BASE, SEAL AND HEADGEAR. THE MASK SEAL IS INSERTED INTO THE MASK BASE AND IS SECURED WITH GLUE. THE RT040 MASK IS SUBJECTED TO POST-PRODUCTION TESTS TO ENSURE THE SEAL ON THE MASK CAN WITHSTAND A REMOVAL FORCE GREATER THAN 100N. WE ARE UNABLE TO DETERMINE THE ROOT CAUSE OF THE SEPARATION. VISUAL INSPECTION OF THE RETURNED DEVICE SHOWED THAT THE MASK WAS ASSEMBLED AND GLUED PROPERLY. THIS SUGGESTS THAT THE SEPARATION OCCURRED POST-PRODUCTION, POSSIBLY DUE TO THE DRYING OUT OR SEPARATION OF THE GLUE CAUSED BY ADVERSE CONDITIONS DURING TRANSPORT OR STORAGE. (B)(4).
A HOSPITAL IN (B)(6) REPORTED VIA OUR DISTRIBUTOR THAT THE SILICONE SEAL OF AN RT040S FULL FACE MASK WAS FOUND SEPARATED FROM THE MASK BASE, CAUSING IT TO LEAK. THIS WAS NOTICED DURING USE ON A PATIENT. NO PATIENT CONSEQUENCE WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | FULL FACE MASK | MNT | MNT | FISHER & PAYKEL HEALTHCARE LIMITED | RT040S | 090807 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |