FDA Adverse Event
Malfunction
Summary report: N
ANIMAS INSULIN INFUSION PUMP
MDR report key: 2244059
·
Received September 12, 2011
Report
- Report Number
- 2531779-2011-06723
- Event Type
- Malfunction
- Date Received
- September 12, 2011
- Report Date
- August 16, 2011
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K042873
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE PUMP HAS BEEN RETURNED TO ANIMAS FOR EVALUATION. THE FOLLOWING INFORMATION WAS FOUND WITH THE SUBJECT PUMP : THE "BOLUS, BACKLIGHT, & UP" BUTTONS ARE NOT RESPONDING. PRODUCT ANALYSIS REMOVED THE KEYPAD AND FOUND ADHESIVE UNDER THE CONTACTS AND REMOVED THE "BOLUS" BUTTON AND FOUND CONTAMINATED CONTACT. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE.
Description of Event or Problem · 1
THE PATIENT CLAIMED THAT THE BOLUS AND UP BUTTONS REQUIRED SEVERAL PRESSES TO ACTIVATE THE DESIRED PUMP FUNCTIONS. THE KEYPAD IS REPORTEDLY INTACT .THERE WAS NO ADVERSE EVENT ASSOCIATE WITH THIS COMPLAINT. THE PUMP WAS REPLACED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | ANIMAS INSULIN INFUSION PUMP | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION | ANM 2020 INSULIN INFUSION PUMP |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 20 YR |