FDA Adverse Event Malfunction Summary report: N

UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM

MDR report key: 2243971 · Received August 19, 2011

Report

Report Number
2122870-2011-02951
Event Type
Malfunction
Date Received
August 19, 2011
Date of Event
August 25, 2008
Report Date
August 26, 2008
Manufacturer
BECKMAN COULTER, INC.
Product Code
JJE
PMA / PMN Number
K023764
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

SERVICE WAS NOT DISPATCHED AS THE CUSTOMER DID NOT QUESTION SYSTEM PERFORMANCE. THE PLASMA SAMPLE WAS COLLECTED IN A LITHIUM HEPARIN TUBE WITHOUT A GEL SEPARATOR. QUALITY CONTROL (QC) WAS WITHIN SPECIFICATIONS PRIOR TO THE EVENT. TESTING AT BECKMAN COULTER CUSTOMER PRODUCT LINE SUPPORT (CPLS) LAB CONFIRMED THE EXISTENCE OF A PT SOURCE INTERFERENT FOR THE SAMPLE RECEIVED FROM THE CUSTOMER. THE INTERFERENCE LIKELY RELATES TO ALKALINE PHOSPHATASE. THIS INTERFERING COMPOUND CAUSES REPRODUCIBLE FALSE POSITIVE RESULTS, BUT IS DISTINCT FROM HETEROPHILE ANTIBODIES. PRODUCT LABELING: FOR ASSAYS EMPLOYING MOUSE ANTIBODIES, THE POSSIBILITY EXISTS FOR INTERFERENCE BY HUMAN ANTI-MOUSE ANTIBODIES (HAMA) IN THE SAMPLE. HUMAN ANTI-MOUSE ANTIBODIES MAY BE PRESENT IN SAMPLES FROM PTS WHO HAVE RECEIVED IMMUNOTHERAPY OR DIAGNOSTIC PROCEDURES UTILIZING MONOCLONAL ANTIBODIES OR IN INDIVIDUALS WHO HAVE BEEN REGULARLY EXPOSED TO ANIMALS. ADDITIONALLY, OTHER HETEROPHILE ANTIBODIES, SUCH AS HUMAN ANTI-GOAT ANTIBODIES, MAY BE PRESENT IN PT SAMPLES. THE ACCESS ACCUTNI RESULTS SHOULD BE INTERPRETED IN LIGHT OF THE TOTAL CLINICAL PRESENTATION OF THE PT, INCLUDING: SYMPTOMS, CLINICAL HISTORY, CLINICAL EXAMINATION, ELECTROCARDIOGRAM (ECG), DATA FROM ADD'L TESTS, AND OTHER APPROPRIATE INFO. MEDICAL DECISIONS SHOULD NOT BE BASED ON A SINGLE ACCUTNI DETERMINATION AT ONE TIME POINT. THIS REPORTABLE EVENT WAS IDENTIFIED DURING A RETROSPECTIVE REVIEW OF PRODUCT COMPLAINTS CONDUCTED FROM (B)(6) 2008 THROUGH (B)(6) 2010 FOR ADD'L REPORTABLE EVENTS.

Description of Event or Problem · 1

THE CUSTOMER REPORTED ELEVATED TROPONIN I (ACCUTNI) RESULTS FOR ONE PT, INVOLVING UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM. THE RESULTS WERE REPRODUCIBLE AND DISCORDANT TO AN ALTERNATE METHODOLOGY. THE ELEVATED RESULTS WERE RELEASED OUT OF THE LAB. THERE HAS BEEN NO REPORT OF PT INJURY OR CHANGE IN PT TREATMENT ASSOCIATED WITH THIS EVENT. THE CUSTOMER SUPPLIED BECKMAN COULTER, INC. WITH THE PT SAMPLES FOR FURTHER ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 UNICEL DXI 600 ACCESS IMMUNOASSAY SYSTEM JJE BECKMAN COULTER, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1 78 YR ACCESS ACCUTNI