I-STAT EG7+ CARTRIDGE
Report
- Report Number
- 2245578-2025-00076
- Event Type
- Malfunction
- Date Received
- July 8, 2025
- Date of Event
- June 25, 2025
- Report Date
- July 22, 2025
- Manufacturer
- ABBOTT POINT OF CARE INC.
- Product Code
- CHL
- PMA / PMN Number
- K940918
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IN
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
- Health Professional
- Yes
Narratives
APOC INCIDENT # (B)(4) APOC LABELING WILL BE EVALUATED DURING THE INVESTIGATION AS PERTAINING TO THE EVENT.
APOC INCIDENT: # (B)(4). THE INVESTIGATION WAS COMPLETED ON 17-JUL-2025. A REVIEW OF THE DEVICE HISTORY RECORD (DHR) CONFIRMED THE CARTRIDGE LOT MET FINISHED GOODS (FG) RELEASE CRITERIA. RETAINED CARTRIDGE TESTING MET THE ACCEPTANCE CRITERIA OUTLINED IN APPENDIX 1 OF Q04.01.003 REV. AO, PRODUCT COMPLAINT LEVEL 2 AND LEVEL 3 INVESTIGATION PROCEDURE. NO DEFICIENCY HAS BEEN IDENTIFIED FOR EG7+ CARTRIDGE LOT N25093.
ON (B)(6) 2025, ABBOTT POINT OF CARE WAS CONTACTED BY A CUSTOMER REGARDING I-STAT EG7+ CARTRIDGES THAT YIELDED A SUSPECTED DESCREPANT PO2 RESULTS ON A PATIENT. THERE WAS NO PATIENT INFORMATION AVAILABLE AT THE TIME OF THIS REPORT. NOR DETAILS SURROUNDING THE EVENT. RETURN PRODUCT IS AVAILABLE FOR INVESTIGATION. DATE: (B)(6)2025, (B)(6)2025 , (B)(6)2025, (B)(6)2025 . TIME: 10:39 , 11:13, 12:01 , 12:06, PH: 7.708 , 7.44, 7.382 , 7.092, PCO [2] (MMHG): 10.8 , 28.1, 36.8 , 47.6, PO2 (MMHG) : 183 , 167 , 23 , 173, BECF (MMOL/L) : -7, -5, -3, -15, HCO [3] (MMOL/L) : 13.1, 19.5, 21.9 , 14.5, TCO [2] (MMOL/L) : 13 , 20 , 23, 16, SO [2] (%) : 100 , 100, 40, 99, NA (MMOL/L) : 135, 136 , 134, 128, K (MMOL/L) : 4.1 , 3.1 , 3.3, >9.0, ICA (MMOL/L) : 0.87, 1.05 , 0.92, *** HCT (%): 37, 34 , 33, 33, HB VIA HCT (G/DL): 12.6 , 11.6 , 11.3, 11.2. AT THIS TIME THERE IS NO REASON TO SUSPECT A MALFUNCTION EXISTS. THE REPORTING DECISION WAS BASED ON INFORMATION AVAILABLE THAT SUGGESTS THE PRODUCT WAS NOT PERFORMING WITHIN THE VARIABILITY OF THE ASSAY. THE INVESTIGATION IS UNDERWAY.
NA.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 388627 | I-STAT EG7+ CARTRIDGE | EG7+ CARTRIDGE | CHL | ABBOTT POINT OF CARE INC. | NA | N25093 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA | Unknown |