FDA Adverse Event Malfunction Summary report: N

INRATIO

MDR report key: 2243681 · Received July 28, 2011

Report

Report Number
2027969-2011-01661
Event Type
Malfunction
Date Received
July 28, 2011
Date of Event
June 24, 2011
Report Date
July 28, 2011
Manufacturer
ALERE SAN DIEGO, INC.
Product Code
GJS
PMA / PMN Number
021923
Removal / Correction Number
NI
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KY, US
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

DISCREPANT RESULTS (ACCURACY) COMPARISON OF INRATIO TEST WITH LAB RESULTS PROVIDED BY END-USER AT TIME COMPLAINT WAS FILED. INRATIO: 4.0, REFERENCE: 3.0, MEAN: 3.50, CONFIDENCE LIMITS: 2.0-5.0. INRATIO: 4.0, 2.9, 3.45, 2.0-5.0. ANALYSIS OF CUSTOMER'S DATA REVEALED THAT BOTH INRATIO AND REFENCE TEST RESULT COMPARISONS MEET ACCURACY CRITERIA. CUSTOMER'S RESULTS ARE NOT CONSIDERED DISCREPANT WITHIN THE CONTEXT OF THE DOCUMENTED VARIABILITY FOR INR TESTING. NO PRODUCT IS EXPECTED TO BE RETURNED. NO FURTHER INVESTIGATION IS REQUIRED. (B)(4). ACTION THRESHOLD (B)(4) WAS REACHED. STRIP LOT # 243104 HAS STRIP CODE Z45BU WITH BRICK LOT # 246544, WHICH WAS ASSIGNED AND PACKAGED INTO TWO STRIP LOTS (243104 AND 251117). (B)(4). DUE TO THIS LOW OCCURRENCE RATE, BELOW THE TOTAL COMPLAINT THRESHOLD OF (B)(4) FOR CORRECTIVE ACTION, NO FURTHER ACTION IS REQUIRED. THIS ISSUE WILL BE SUBJECT TO TRACKING AND TRENDING. NO CORRECTIVE ACTION REQUIRED AT THIS TIME AS NO PRODUCT DEFICIENCY WAS ESTABLISHED.

Description of Event or Problem · 1

CALLER ALLEGED DISCREPANT RESULTS COMPARED WITH THE LAB AND ANOTHER POINT-OF-CARE (POC) DEVICE. RESULTS AS FOLLOWS: DATE: (B)(6) 2011, INRATIO: 4.0, LAB: 2.9, POC: 3.0. PT'S THERAPEUTIC RANGE: 2.0-3.0 INR.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 INRATIO PROTHROMBIN TIME TEST GJS ALERE SAN DIEGO, INC. 100071 243104

Patients

Seq Age Sex Outcome Treatment
1 NI