FDA Adverse Event Malfunction Summary report: N

2800

MDR report key: 2243659 · Received August 24, 2011

Report

Report Number
1720753-2011-21316
Event Type
Malfunction
Date Received
August 24, 2011
Date of Event
August 15, 2011
Report Date
August 24, 2011
Manufacturer
GE OEC MEDICAL SYSTEMS (SLC)
Product Code
JAA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

A GE SERVICE REP PERFORMED AN ONSITE INVESTIGATION. THE SERVICE REP REPLACED THE POWER SUPPLY 4 AND SET THE 5 VDC TO 5.15 VDC. THE SYSTEM WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK IN SERVICE.

Description of Event or Problem · 1

THE CUSTOMER REPORTED THAT THE SYSTEM WOULD NOT PERFORM FLUOROSCOPIC X-RAY. THIS HAPPENED DURING A PROCEDURE. NO PT INJURY WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 2800 FLUOROSCOPIC X-RAY JAA GE OEC MEDICAL SYSTEMS (SLC) 2800

Patients

Seq Age Sex Outcome Treatment
1