FDA Adverse Event
Malfunction
Summary report: N
9900
MDR report key: 2243623
·
Received August 24, 2011
Report
- Report Number
- 1720753-2011-21350
- Event Type
- Malfunction
- Date Received
- August 24, 2011
- Date of Event
- August 16, 2011
- Report Date
- August 24, 2011
- Manufacturer
- GE OEC MEDICAL SYSTEMS (SLC)
- Product Code
- JAA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
A GE SVC REP PERFORMED AN ON SITE INVESTIGATION. THE REPORTED ISSUE COULD NOT BE DUPLICATED. THE BOARDS AND CONNECTORS WERE RESEATED DURING THE SVC CALL. THE SYS WAS TESTED AND FOUND TO BE WORKING AS INTENDED AND PUT BACK INTO SVC.
Description of Event or Problem · 1
THE CUSTOMER REPORTED THAT THE SYS WOULD NOT BOOT UP. NO PT INJURY WAS REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | 9900 | FLUOROSCOPIC X-RAY | JAA | GE OEC MEDICAL SYSTEMS (SLC) | 9900 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |