FDA Adverse Event Malfunction Summary report: N

LIFEPAK CR PLUS DEFIBRILLATOR

MDR report key: 2243617 · Received July 27, 2011

Report

Report Number
3015876-2011-00559
Event Type
Malfunction
Date Received
July 27, 2011
Date of Event
July 1, 2011
Report Date
July 1, 2011
Manufacturer
PHYSIO-CONTROL, INC.
Product Code
MKJ
PMA / PMN Number
K011144
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
VA, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). PHYSIO-CONTROL EVALUATED THE DEVICE AND VERIFIED THE REPORTED FAILURE. PHYSIO DETERMINED THAT THE CAUSE OF THE REPORTED FAILURE WAS DUE TO A FAILURE OF AN INTEGRATED CIRCUIT CHIP, DESIGNATOR U29 FROM THE DIGITAL PCB ASSEMBLY. THE CUSTOMER WAS PROVIDED WITH A REPLACEMENT DEVICE.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE DEVICE DISPLAYED ALL THREE (ATTENTION, CHARGE-PAK AND WRENCH) ICONS AND FAILED TO POWER ON. THERE WAS NO PATIENT USE ASSOCIATED WITH THE REPORTED FAILURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 LIFEPAK CR PLUS DEFIBRILLATOR MKJ PHYSIO-CONTROL, INC. CRPLUS NA

Patients

Seq Age Sex Outcome Treatment
1 NA