FDA Adverse Event Malfunction Summary report: N

VIRTUOSO VR

MDR report key: 2243581 · Received September 12, 2011

Report

Report Number
6000094-2011-01833
Event Type
Malfunction
Date Received
September 12, 2011
Date of Event
July 26, 2011
Manufacturer
MEDTRONIC S.A.
Product Code
LWS
PMA / PMN Number
ASKU
Removal / Correction Number
ASKU
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S.

Additional Manufacturer Narrative · 1

THIS EVENT OCCURRED OUTSIDE THE US. ALL INFORMATION PROVIDED IS INCLUDED IN THIS REPORT. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. PATIENT INFORMATION IS NOT GENERALLY AVAILABLE DUE TO CONFIDENTIALITY CONCERNS. THIS DEVICE MODEL NUMBER IS NOT APPROVED FOR DISTRIBUTION IN THE UNITED STATES, HOWEVER, IT IS SAME/SIMILAR TO A DEVICE MARKETED IN THE U.S.

Description of Event or Problem · 1

IT WAS REPORTED THAT WHILE WASHING DISHES, THE PATIENT EXPERIENCED NUMBNESS IN THE HAND AND APPROXIMATELY TWO MINUTES LATER RECEIVED THERAPY. IT WAS FURTHER REPORTED THAT PHYSICIAN REVIEW OF THE EPISODE DETERMINED THERAPY WAS DUE TO ELECTRICAL INTERFERENCE. FURTHER INVESTIGATION BY THE FACILITY REVEALED STRAY VOLTAGE FROM EXTERNAL EQUIPMENT WAS PRESENT AT THE TIME OF THE OCCURRENCE. THE DEVICE REMAINS IN USE. NO FURTHER PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Description of Event or Problem · 1

ASKU

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 VIRTUOSO VR IMPLANTABLE PACEMAKER/CARDIO/DEFIB LWS MEDTRONIC S.A. D164VWC ASKU

Patients

Seq Age Sex Outcome Treatment
1 17 YR Life Threatening| O